0595. Quantitating granulocyte reactive oxygen species production by flow cytometry in a clinical setting
Intensive Care Medicine Experimental volume 2, Article number: P38 (2014)
Introduction
Oxidative stress is an important part of a wide range of pathologies and therefore an interesting parameter to determine. However, the detection of reactive oxygen species (ROS), is not straightforward. Sample heterogeneity, delayed analysis and sample preparation, reduces specificity and sensitivity due to probe activation by light, air, probe leakage from cells or cellular activation by sample manipulation. Hence, samples should be processed minimally and analysed immediately if possible. Since clinical research can be subject to uncontrollable timetables and only few departments have access to a dedicated laboratory, the quantification of ROS is challenging.
Objectives
For the purpose of a randomized-controlled trial involving Coronary Artery Bypass Graft surgery (CABG), we sought to establish a flow cytometric assay to detect ROS production in granulocytes (PMN) that is easy to perform, does not require a specialised laboratory on-site and is relatively insensitive to delay between sample preparation and analysis. The fluorescent probe applied is CellROX® Green (CrX; Life Technologies, Eugene, Oregon, USA), which becomes fluorescent after oxidation and subsequently binds to DNA.
Methods
Whole blood samples were collected from healthy volunteers and from patients before (baseline) and during CABG surgery (after 1 hour of cardiopulmonary bypass, CPB). 30 µl whole blood was incubated for 1 hour with 10µM CrX immediately after sampling. To compensate for the surgery induced PMN count increase, 1 hour CPB samples were incubated with 10, 20 and 30µM CrX. Following incubation, erythrocytes were lysed by hypotonic shock and the sample was washed three times with Phosphate Buffered Saline (PBS). Finally, leukocytes were stored in 200µl of PBS at room temperature, in the dark, until analysis. Positive and negative controls were obtained by additional incubation with phorbol-12-myristate-13-acetate and N-acetyl-cysteine. Mean Fluorescence Intensity (MFI) of the PMN population, which is proportional to the ROS produced, was measured in the FL1 channel of a BD FACSCalibur as soon as possible after incubation of the second CABG sample (~5 hours). PMN were identified based on their forward and sideward scatter characteristics to avoid fluorescence compensation corrections and aspecific activation of neutrophils by fluorescent cell-specific antibodies.
Results
In line with our expectations, little production of ROS was demonstrated in healthy volunteers (n=4) and baseline samples from CABG surgery patients, while a marked increase was found after one hour of CPB (n=8, figure 1). PMA induces a NAC-inhibitable increase of MFI in CABG baseline samples. Interestingly, we were also able to detect what appears as separate ROS producing neutrophil populations, which warrants further investigation.
Conclusions
Our current method is easy to perform, implementable into clinical research and is able to detect different ROS responses.
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Smit, B., de Waard, M., Smulders, Y. et al. 0595. Quantitating granulocyte reactive oxygen species production by flow cytometry in a clinical setting. ICMx 2 (Suppl 1), P38 (2014). https://doi.org/10.1186/2197-425X-2-S1-P38
Published:
DOI: https://doi.org/10.1186/2197-425X-2-S1-P38