Skip to main content

Volume 3 Supplement 1

ESICM LIVES 2015

  • Poster presentation
  • Open access
  • Published:

Reliability of diaphragmatic ultrasonography to detect diaphragm dysfunction in critically ill patients

Introduction

The gold standard to diagnose diaphragmatic dysfunction (DD) is the measurement of the intra-thoracic depression is response to a bilateral stimulation of the phrenic nerves (Ptr,stim). This technique is costly, requires expertise and is not widely available at the bedside. On the opposite, ultrasonography is easy to perform but has not been compared.

Objectives

To validate diaphragm ultrasonography as a tool to detect diaphragm dysfunction in mechanically ventilated patients. Ptr,stim was used as the gold standard.

Methods

This monocentric prospective study was conducted in a 16-beds medical ICU. DD was assessed by two independent observers at three key time points:

1) during the first 24h of controlled ventilation (CV),

2) as soon as patients could tolerate pressure support ventilation (PSV) and

3) the day of the first spontaneous breathing trial (SBT). Ptr,stim was measured in response to bilateral anterior magnetic stimulation of the phrenic nerves. M-mode ultrasonography was used measure

1) the muscle thickening fraction (TF, (defined as the difference between inspiratory and expiratory muscle thickness divided by the expiratory thickness) of the right hemidiaphragm,

2) the maximal right diaphragmatic excursion was also recorded. A Ptr, stim < -11 cmH2O defined DD.

Results

Seventy-three critically ill patients were investigated (136 measurements: 54 under CV, 32 under PSV and 40 the day of the SBT). DD was present in 79%, 85% and 60% of the patients under CV, PSV and SBT respectively. Taken together, the measurements performed under PSV and the day of the SBT showed a significant correlation between Ptr,stim and TFd (R2 = 0.79, p < 0.001). A TF < 29% had a sensitivity and specificity of 85% and 95% to detect DD (area under the receiver operating characteristics curve 0.93).

There were no significant correlation between Ptr,stim and

  1. 1)

    diaphragm thickness

  2. 2)

    the maximal excursion of the diaphragm.

Conclusions

Diaphragm ultrasonography is a reliable method to identify DD in mechanically ventilated patients with an active inspiration. A diaphragm TF < 29% is a strong predictor of DD.

Grants

Martin DRES was supported by Assistance Publique Hôpitaux de Paris

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Dres, M., Dubé, BP., Mayaux, J. et al. Reliability of diaphragmatic ultrasonography to detect diaphragm dysfunction in critically ill patients. ICMx 3 (Suppl 1), A452 (2015). https://doi.org/10.1186/2197-425X-3-S1-A452

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2197-425X-3-S1-A452

Keywords