Temperature Monitoring During Ecmo: An in Vitro Study

Introduction

The need of heat exchanger in the ECMO circuit is controversial. Moreover, how to monitor patient central temperature during extracorporeal support is still not clear, but potentially useful for the detection of “unexpressed” fever, eventually related to septic complications.

Objectives

We conducted two in vitro experiments to estimate ECMO heat dispersion and obtain clinical information regarding patient central temperature during extracorporeal support.

Methods

Experiment A. We analyzed heat dispersion of an ECMO circuit (Maquet Rotaflow PLS System) at 36 combinations of blood flow (BF 1-3-5 L/min), gas flow (GF 0-5-10 L/min) and set temperature (T set 36-37-38-39 °C). in any condition, heat dispersion was considered equal to the power (Watts) generated by the heat exchanger at the steady state, defined as stable temperature throughout the circuit.

Experiment B. By a reservoir bag, we connected two circuits (pump+oxygenator+heat exchanger), one simulating the patient and the other the ECMO circuit. Patient and ECMO circuit T set ranged 36-39 °C and 35-39 °C, respectively, for overall 63 conditions; ΔT was patient - ECMO circuit T set difference. the power generated by the two heat exchangers (Watts, W) was recorded at constant patient BF (5 L/min) and ECMO BF and GF (3 L/min each).

Results

A) Overall, BF was 3.0 ± 1.7 L/min, GF 5.0 ± 4.1 L/min, pump rate 1299 ± 615 rpm, T set 37.5 ± 1.1 °C and circuit heat dispersion 58 ± 12 W (range 37.5-90), corresponding to a supposed patient metabolic consumption of 1196 ± 246 Kcal/die (range 774-1857) in case of heat exchanger absence or inactivation. It was affected by GF and T (p < 0.001), but not by BF (p = 0.9) (Figure 1).

figure 1

Figure 1

B) For ΔT > 0 heat exchanger did not generate energy, while for ΔT < 0 it supplied energy proportional to ΔT (p < 0.001). For ΔT = 0, heat exchanger power was 49 ± 7 W (range 40-55) (Figure 2).

figure 2

Figure 2

Conclusions

ECMO heat dispersion depends on GF and temperature, but not on BF. Therefore, heat exchanger should be considered also during low BF/high GF ECCO2R. T set on heat exchanger is well matched with patient central temperature when its power is in the 40-55 W range, whereas lower power values may be associated to patient “unexpressed” fever.

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Mojoli, F., Bianzina, S., Caneva, L. et al. Temperature Monitoring During Ecmo: An in Vitro Study. ICMx 3 (Suppl 1), A506 (2015). https://doi.org/10.1186/2197-425X-3-S1-A506

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2197-425X-3-S1-A506

Keywords