Skip to main content

Table 1 Summary of in vitro hemolysis testing

From: An extracorporeal carbon dioxide removal (ECCO2R) device operating at hemodialysis blood flow rates

Test device

Blood flow

TIH

NIH

(mL/min)

(g/100 min)

(g/100 L)

ULFED (standard circuit)

250

0.190

±0.041

0.775

±0.186*

ULFED (dialysis configuration)

250

0.386

±0.010*

1.551

±0.025*

Minimax

1250

0.123

±0.013

0.105

±0.012*