Skip to main content
Fig. 4 | Intensive Care Medicine Experimental

Fig. 4

From: Bias flow rate and ventilation efficiency during adult high-frequency oscillatory ventilation: a lung model study

Fig. 4

Relationship between actual stroke volume (aSV) and alveolar ventilation (\( \dot{\mathrm{V}} \)A). Legend: Markers indicate individual data of aSV and \( \dot{\mathrm{V}} \)A. Dotted curves are power approximations. The perfect powers (coefficient of correlation: R, P value) are as follows: BF = 10: 1.128 (R = 0.996, P < 0.001); BF = 20: 1.528 (R = 0.995, P < 0.001); BF = 30: 1.631 (R = 0.992, P < 0.001); and BF = 40: 1.664 (R = 0.992, P < 0.001)