Fig. 6 | Intensive Care Medicine Experimental

Fig. 6

From: Bias flow rate and ventilation efficiency during adult high-frequency oscillatory ventilation: a lung model study

Fig. 6

Alveolar ventilation (\( \dot{\mathrm{V}} \)A) measured with targeted actual stroke volume (aSV) of 80, 100, and 120 mL. Legend: Bar graph indicates mean \( \dot{\mathrm{V}} \)A (n = 5), and vertical bar indicates standard deviation. The results of statistical significance test are as follows: BF = 10 vs BF = 20–60: P < 0.001 with all aSV; BF = 20 vs BF = 30–60: P < 0.001 with all aSV; BF = 30 vs BF = 40: ns with aSV = 80 and 120, P < 0.05 with aSV = 100; BF = 30 vs BF = 50: P < 0.01 with aSV = 80 and 120, P < 0.001 with aSV = 100; BF = 30 vs BF = 60: P < 0.001 with all aSV; BF = 40 vs BF = 50: ns with all aSV; BF = 40 vs BF = 60: P < 0.05 with aSV = 80 and 100, P < 0.001 with aSV = 120; and BF = 50 vs BF = 60: ns with all aSV