Skip to main content
Fig. 8 | Intensive Care Medicine Experimental

Fig. 8

From: Bias flow rate and ventilation efficiency during adult high-frequency oscillatory ventilation: a lung model study

Fig. 8

Relationship between actual minute volume divided by bias flow (a\( \dot{\mathrm{V}} \)E/BF) and alveolar ventilation (\( \dot{\mathrm{V}} \)A). Legend: Dotted curves are the exponential approximations. The coefficient of correlations (R) and P values are as follows: aSV = 80: R = − 0.982, P < 0.001; SV = 100: R = − 0.991, P < 0.001; and aSV = 120: R = − 0.992, P < 0.001. The Y-intercepts are 7.54 (aSV = 80), 11.46 (aSV = 100), and 15.31 (aSV = 120)

Back to article page