Skip to main content

Table 1 In vitro hemolysis of the LF-PAL

From: Bench Validation of a Compact Low-Flow CO2 Removal Device

Device Flow Rate NIH TIH
[mL/min] [g/100 L] [g/100 min]
LF-PAL (0.65 m2) 500 0.158 ± 0.034 0.080 ± 0.017
Pedimag control 1500 0.029 ± 0.003 0.043 ± 0.004
BP-50 control 1250 0.105 ± 0.012 0.123 ± 0.013
  1. Statistically significant (p < 0.05) compared to other devices
  2. Statistically significant (p < 0.05) compared to other devices