Table 2 Comparison of translaryngeal Tracheostomy Needle Introducer versus standard needle puncture technique among all operators

From: Translaryngeal Tracheostomy Needle Introducer: a simple device to improve safety and reduce complications during Fantoni’s translaryngeal tracheostomy procedure: trial on human cadavers

Attempts by all operators

(N = 58)

Average ± SD

Min

Max

Median

25th percentile

75th percentile

p value

Attempts with standard method (N)

1.91 ± 1.34

1

7

1

1

2

p < 0.001

Attempts with tTNI (N)

1.19 ± 0.5

1

3

1

1

1

Time required for standard method (s)

36.8 ± 51.6

3

300

20

13

35.75

p < 0.001

Time required for tTNI (s)

13.14 ± 15.57

2

92

8

7

11

Precision of first attempt* with standard method (°)

18.87 ± 25.38

0

120

15

0

30

p < 0.005

Precision of first attempt* with tTNI (°)

7.5 ± 12.95

0

60

0

0

15

  1. Data analysed using the Wilcoxon sign test
  2. *Endoscopic reference: degrees from the anterior tracheal midline with the target ranging from − 30° to + 30°