Fig. 5 | Intensive Care Medicine Experimental

Fig. 5

From: In vitro characterization of PrismaLung+: a novel ECCO2R device

Fig. 5

Comparison of CO2 removal performance using the infusion and blood gas analysis methods. Assessed across all tested devices and parameter settings at 37 °C. a Bland–Altman analysis to demonstrate the relationship between the BGA method and the CO2 input method. Mean = 4.2 mL/min, SD = 4.4 mL/min, Limits of Agreement (95%) = − 4.4 mL/min, 12.9 mL/min. b Linear regression analysis comparing the two test methods: BGA method and CO2 input method. Slope not significantly different from 1.0. BGA, blood gas analysis; SD, standard deviation