From: In vitro characterization of PrismaLung+: a novel ECCO2R device
 | Mean normalized JCO2 removal rate, mL/min (SD) | p value | ||||
---|---|---|---|---|---|---|
Set QB, mL/min | Set pinCO2, mmHg | PrismaLung+ | PrismaLung | A.L.ONE | PrismaLung+ vs. PrismaLung | PrismaLung+ vs. A.L.ONE |
200 | 45 | 54 (3.8) | 35 (2.5) | 54 (0.7) | < 0.001 | 0.62 |
300 | 45 | 73 (4.0) | 44 (2.5) | 72 (1.9) | < 0.001 | 0.59 |
450 | 45 | 93 (4.0) | 54 (3.2) | 92 (2.7) | < 0.001 | 0.63 |
200 | 60 | 65 (5.0) | 43 (1.9) | 64 (1.4) | < 0.001 | 0.70 |
300 | 60 | 86 (5.7) | 55 (3.0) | 85 (2.8) | < 0.001 | 0.66 |
450 | 60 | 111 (7.8) | 66 (3.4) | 110 (4.3) | < 0.001 | 0.76 |