From: A mock circulation loop to test extracorporeal CO2 elimination setups
Simulated cannula flow [mL/min] | Recirculation channel flow [mL/min] | Membrane flow [mL/min] | Recirculation ratio [%] | \( {\dot{\boldsymbol{V}}}_{{\mathbf{CO}}_{\mathbf{2}}} \) [mL/min] | p |
---|---|---|---|---|---|
910 | Closed | 910 | 66.67 ± 0.69 | p < 0.05 | |
900 | 500 | 1400 | 35.7 | 69.63 ± 0.73 | |
500 | Closed | 500 | 58.31 ± 0.61 | ns | |
500 | 280 | 780 | 35.9 | 57.97 ± 0.4 | |
300 | Closed | 300 | 48.66 ± 0.46 | p < 0.05 | |
300 | 240 | 540 | 44.4 | 49.93 ± 0.43 | |
200 | Closed | 200 | 42.25 ± 0.71 | ns | |
200 | 140 | 340 | 41.2 | 42.32 ± 0.99 | |
100 | Closed | 100 | 28.76 ± 1.22 | p < 0.05 | |
100 | 80 | 180 | 44.4 | 30.93 ± 0.62 | |
330 | Closed | 330 | 55.38 ± 0.88 | ns | |
330 | 1140 | 1470 | 77.6 | 54.45 ± 1.26 | |
180 | Closed | 180 | 45.16 ± 0.96 | p < 0.05 | |
180 | 1310 | 1490 | 87.9 | 41.24 ± 0.43 |