Table 3 Comparison of CO2 removal rates of the pediatric oxygenator Quadrox iD for different blood flow (BF) and sweep gas flows (\( {\dot{V}}_{\mathrm{Sweep}} \))

From: A mock circulation loop to test extracorporeal CO2 elimination setups

\( {\dot{V}}_{\mathrm{Sweep}} \)

This study (BF 0.6 L/min)

Sun et al. (BF 0.5 L/min) [12]

1 L/min

41 mL/min

39 mL/min

2 L/min

63 mL/min

61 mL/min

4 L/min

85 mL/min

92 mL/min