Skip to main content

Table 1 Comparison between inspiration synchronized and continuous nebulizer at different placement with quiet vs distressed breathing

From: In vitro comparison between inspiration synchronized and continuous vibrating mesh nebulizer during trans-nasal aerosol delivery

Nebulizer position

Breathing pattern

HFNC gas flow rate (L/min)

Inhaled dose (%)

Inspiration synchronized nebulizer

Continuous nebulizer

Inlet of humidifier

Quiet (500 mL/15 bpm/1:2/22.5 L/min)

5

19.8 ± .8

19.5 ± .3

10

27.6 ± 1.2

19.8 ± .8

20

9.6 ± .3

16.5 ± .8

40

3.5 ± .1

8.8 ± 1.0

60

1.1 ± .04

6.3 ± .2

Distressed (700 mL/30 bpm/1:1/42 L/min)

5

21.0 ± .6

12.9 ± 1.6

10

23.9 ± 1.3

17.5 ± 1.0

20

31.2 ± 1.3

17.8 ± .2

40

9.1 ± .2

16.5 ± .3

60

2.3 ± .2

9.8 ± .7

Close to patient

Quiet (500 mL/15 bpm/1:2/22.5 L/min)

5

13.4 ± .3

24.1 ± .8

10

24.1 ± 1.0

15.8 ± .6

20

18.0 ± 1.8

10.0 ± .4

40

7.8 ± .6

3.9 ± .3

60

4.7 ± .4

2.0 ± .1

Distressed (700 mL/30 bpm/1:1/42 L/min)

5

12.5 ± 1.4

19.0 ± 1.0

10

24.2 ± .9

17.8 ± .5

20

23.7 ± .7

13.6 ± .2

40

12.0 ± .3

8.3 ± .6

60

6.9 ± .1

3.8 ± .1