Audit of operational noise at the leicester royal infirmary aicu
Intensive Care Medicine Experimental volume 3, Article number: A146 (2015)
Introduction
Ambient noise and its impact on sleep is an important factor in the development of delirium[1]. Noise is a problem on intensive care units, being one of the main environmental stress inducing factors amongst patients[2]. A key source of noise is the signalling of alarms.
As part of plans to decrease noise, we investigated the number and appropriateness of alarms sounded overnight. An initial audit demonstrated that over 53% of alarms were inappropriate, with two recurring sources being due to sampling from indwelling lines and physiological variables falling outside of a pre-programmed range.
Following this we implemented a brief educational programme, highlighting the importance of setting appropriate alarm limits and muting prior to sampling from lines. We then re-audited to determine if improvement had been made.
Objectives
Improvement in the frequency of inappropriately sounded alarms at night following intervention.
Audit standard: 90% of alarms being deemed as appropriate.
Methods
Data was collected prospectively by nursing staff on the intensive care unit. Nurses were allocated to collect data on a bed-space other than their own to reduce bias. Data collected included the source and response to the alarm. Data was reviewed by the authors to determine if the alarms and response were appropriate.
Results
Data was collected over 7 nights with the total number of recorded alarms being 200. Alarm sources included patient monitoring, infusion pumps and ventilators amongst others. 39% of the alarms were deemed to be inappropriate with 7.7% of these being due to sampling from arterial lines. Other frequently occurring alarms included physiological variables and minute/tidal volumes falling outside of pre-determined ranges (34.6% and 15.4% respectively).
Conclusions
Compared to the first audit, the number of alarms deemed as inappropriate has reduced from 53.9% to 39%. Alarms due to sampling from indwelling lines have reduced from 20.8% to 7.7%. Alarms due to physiological variables falling outside of the pre-determined ranges have increased from 27.1% to 34.6%.
Ensuring quality sleep forms an integral part of a patient's recovery from critical illness and its lack thereof plays an important role in the development of delirium with its subsequent effects on morbidity and mortality. Excess ambient noise due to inappropriate alarms will impact on sleep.
Despite some improvement, we have failed to meet our self-imposed audit standard. We plan to reaffirm the muting of alarms prior to taking blood samples and reinforce the importance of appropriate setting of physiological variable limits at the start of each clinical shift. We plan to implement a daily checklist to ensure this is complete.
References
Schiemann A, Hadzidiakos D, Spies C: Managing ICU delirium. Curr Opin Crit Care. 2011, 12: 131-40.
Cornock MA: Stress and the intensive care patient: perceptions of patients and nurses. J Adv Nurs. 1998, 27 (3): 518-27. 10.1046/j.1365-2648.1998.00555.x. Mar
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Charlton, M., Prosho, K. & O'Neil, D. Audit of operational noise at the leicester royal infirmary aicu. ICMx 3 (Suppl 1), A146 (2015). https://doi.org/10.1186/2197-425X-3-S1-A146
Published:
DOI: https://doi.org/10.1186/2197-425X-3-S1-A146