Table 1 Mean signal intensities (measured in decibels) pre- and post-circuit for the combined speed data (with and without an oxygenator in the circuit)

From: Quantification of perflutren microsphere contrast destruction during transit through an ex vivo extracorporeal membrane oxygenation circuit

 

Pre-circuit (dB)

Post-circuit (dB)

Difference

p value

Oxygenator 150 ml/h

24.6 ± 1.7

18.9 ± 2.4

5.7 ± 3.0

p < 0.0001

Oxygenator 300 ml/h

25.9 ± 1.9

20.8 ± 2.6

5.1 ± 2.1

p < 0.0001

Oxygenator combined

25.3 ± 1.9

19.9 ± 2.6

5.4 ± 2.6

p < 0.0001

No oxygenator 150 ml/h

30.9 ± 1.3

29.6 ± 1.4

1.3 ± 1.5

p < 0.0001

No oxygenator 300 ml/h

30.6 ± 3.1

28.8 ± 2.5

1.8 ± 1.8

p < 0.0001

No oxygenator combined

30.8 ± 2.4

29.2 ± 2.0

1.6 ± 1.7

p < 0.0001