Skip to main content

Table 99 (abstract A1164). Effects on patient-ventilator interaction of pressure support, optimization of ETS, and IntelliSync + .

From: ESICM LIVES 2016: part three

 

PS basal ETS 25

PS basal ETS opti1

PS basal IntelliSync+

PS + 50 ETS opti1

PS + 50 ETS opti2

PS + 50 IntelliSync+

Unassisted breaths (%)

12,4

3,5*

5,6*

12,1

10,8

5,6*

Trigger delay (ms)

238 ± 92

240 ± 89

209 ± 158**

295 ± 122**

265 ± 104

249 ± 117

Cycling delay (ms)

282 ± 315

113 ± 143***

54 ± 152****

304 ± 229

121 ± 134***

43 ± 149****

  1. *p < 0.0001 vs. PS basal - ETS 25, PS + 50 ETS opti1 and PS + 50 - ETS opti2
  2. **p < 0.0005 vs. all the other conditions
  3. ***p < 0.0001 vs. PS basal - ETS 25, PS + 50 ETS opti1, PS basal - Intellisync + and PS + 50 - Intellisync+