Skip to main content

Table 4 Comparison of translaryngeal Tracheostomy Needle Introducer versus standard needle puncture technique among untrained operators

From: Translaryngeal Tracheostomy Needle Introducer: a simple device to improve safety and reduce complications during Fantoni’s translaryngeal tracheostomy procedure: trial on human cadavers

Attempts by untrained operators

(N = 32)

Average ± SD

Min

Max

Median

25th percentile

75th percentile

p value

Attempts with the standard method (N)

1.87 ± 1.53

1

7

1

1

2

p < 0.050

Attempts with the tTNI (N)

1.28 ± 0.5

1

3

1

1

1

Time required for standard method (s)

46.8 ± 65.73

6

300

22.5

17

36.25

p < 0.001

Time required for tTNI (s)

17.21 ± 19.7

4

92

10

7.75

15.25

Precision of the first attempt* with standard method

22.03 ± 30.47

0

120

15

0

30

p < 0.020

Precision of the first attempt* with the tTNI

5.16 ± 12.98

0

60

0

0

0

  1. *Endoscopic reference: degrees from the anterior tracheal midline with the target ranging from − 30° to + 30°