Fig 2

From: Respiratory and metabolic acidosis correction with the ADVanced Organ Support system

Fig 2

Results from blood gas analysis obtained after hemodialysis of swine blood under different treatment settings with ADVOS. Experiments with dialysate pH 10 were performed with the basic concentrate without Na2CO3 (BC-Bic 0). The value “10-P” corresponds to the experiments where CO2 was supplied as long as physiological blood gas values were maintained (pH 7.35–7.45, pCO2 35–45 mmHg, HCO3 22–28 mmol/l) and not only blood pH. Mean ± SD.1, p < 0.05 for CO2 removal between consecutive dialysate pH settings among the same blood flow; 2, p < 0.05 for CO2 removal between consecutive blood flows among the same dialysate pH setting; 3, p < 0.05 for CO2 removal between consecutive concentrate flows among the same blood flow and dialysate pH setting