Table 1 Impact of pure oxygen and air as sweep gases on CO2 removal

From: A mock circulation loop to test extracorporeal CO2 elimination setups

\( {\dot{\boldsymbol{V}}}_{\mathbf{Sweep}} \) [L/min]

Sweep gas

pgCO2 [kPa]

\( {\dot{\boldsymbol{V}}}_{{\mathbf{CO}}_{\mathbf{2}}} \) [mL/min]

p

7

O2

1.41

98.92 ± 2.28

p < 0.05

7

Air

1.26

89.68 ± 4.06

 

6

O2

1.56

94.03 ± 0.97

p < 0.05

6

Air

1.35

81.37 ± 1.07

 

4

O2

2.12

85.05 ± 1.04

p < 0.05

4

Air

1.77

72.00 ± 1.16

 

2

O2

3.09

62.51 ± 0.98

p < 0.05

2

Air

2.74

54.67 ± 0.64

 

1

O2

4.05

40.91 ± 1.41

p < 0.05

1

Air

3.58

36.14 ± 0.57

 

0.7

O2

4.35

30.14 ± 0.62

ns

0.7

Air

4.27

29.61 ± 0.55

 

0.5

O2

4.58

23.29 ± 0.32

p < 0.05

0.5

Air

4.12

20.11 ± 0.4

 

0.3

O2

4.7

13.99 ± 0.25

p < 0.05

0.3

Air

4.3

13.13 ± 0.11