Table 2 Bland–Altman analysis of respiratory system’s static compliance and resistance values calculated by the static and dynamic methods for individual patients

From: A novel method to calculate compliance and airway resistance in ventilated patients

Patient

Epochs

(3174)

Breaths

(78,371)

Compliance mL·cmH2O −1

Resistance cmH2O·s·L−1

Mean value

Bias

LOA

Mean value

Bias

LOA

1

293

5924

35.8

0.4

1.1

21.9

0.4

1.4

2

119

4917

27.5

− 0.6

0.9

15.2

0.2

2.2

3

127

2926

58.6

1.4

0.9

14.3

0.2

0.3

4

385

8632

32.8

− 0.3

1.2

20.5

− 0.3

0.9

5

15

459

59.7

1.6

0.9

16.7

− 0.1

0.6

6

38

660

46.7

1.6

2.3

13.6

− 0.4

1.9

7

399

11,714

39.7

0.2

0.9

15.4

0.1

1.1

8

881

17,953

53.5

0.8

0.8

16.8

− 0.3

0.6

9

72

1718

49.7

1.3

0.8

17.7

-0.1

0.5

10

27

426

41.2

1.4

1.3

17.4

2.1

1.1

11

38

1123

47.5

0.7

1.1

12.2

0.8

0.4

12

26

494

52.5

0.0

0.7

12.0

− 0.1

0.2

13

248

6771

34.8

− 0.8

0.7

18.9

− 1.6

0.8

14

476

14,067

33.8

− 0.5

1.2

15.6

− 0.5

1.6

15

30

587

64.7

0.6

1.3

12.9

0.2

0.3

Median

119

2926

46.7

0.6

0.9

15.6

− 0.1

0.8

IQR

[34,339]

[624,7702]

[35.3, 53.0]

[− 0.2, 1.4]

[0.8, 1.2]

[13.9, 17.5]

[− 0.3, 0.2]

[0.5, 1.2]

  1. Mean value = average value of all measurements used in the Bland–Altman analysis; LOA = 95% limits of agreement; IQR = interquartile range]