Table 3 EIT results PCV (baseline) vs optimized FCV (step 3); values represent median (IQR)

From: Flow-controlled ventilation decreases mechanical power in postoperative ICU patients

 

Optimized FCV

p value

a. Changes in EIT parameters during FCV as compared to PCVa

 Global change in ΔZ (%)

59.4 (34.3–72.1)

 

 Regional change in ΔZ (%)

0.0301

  ROI ventral

39.7 (22.1–49.5)

 

  ROI mid-ventral

50.9 (26.6–66.3)

  ROI mid-dorsal

73.6 (34.3–78.6)

  ROI dorsal

81.1 (52.7–104.7)

 Global change in static compliance (%)

13.4 (8.0–26.7)

 

 Regional change in static compliance (%)

 

0.0172

  ROI ventral

2.4 (− 7.0–19.2)

 

  ROI mid-ventral

15.2 (2.7–24.2)

  ROI mid-dorsal

23.3 (7.1–32.6)

  ROI dorsal

27.5 (19.9–45.2)

Change in global EELI (a.u.)

53 (− 17–100)

0.163

 

PCV

Optimized FCV

p value

b. Absolute EIT parameters reflecting lung and ventilation homogeneity

 GI (%)

43.8 (41.4–45.3)

43.5 (39.7–45.7)

1.000

 RVDi (%)

2.75 (2.28–4.63)

4.23 (3.39–6.11)

0.717

  1. a.u. arbitrary units, EELI end-expiratory lung impedance, EIT electrical impedance tomography, FCV flow-controlled ventilation, GI global inhomogeneity index, PCV pressure-controlled ventilation, ROI region of interest, RVDi regional ventilation delay index
  2. aChanges in ΔZ and static compliance are expressed as percentage change between FCV step 3 and PCV at baseline, since both are expressed in arbitrary units which makes direct comparisons between patients unreliable
  3. 1p value reflects the significant difference between PCV baseline vs. FCV step 3 regarding the distribution of ΔZ among the four ROIs, using a Kruskall Wallis test on the percentage changes from baseline (to account for the fact that ΔZ is measured in arbitrary units). 2p value reflects the significant difference between PCV baseline vs. FCV step 3 regarding the distribution of the change in static compliance among the four ROIs, using a Kruskall Wallis test on the percentage changes from baseline (to account for the fact that ΔZ and thereby also the static compliance is measured in arbitrary units)