Background/Purpose
Medical flow cytometry (FACS) provides diagnostic answers by detecting the presence and concentration of cell populations, and/or by measuring concentrations of cell surface markers expressed on cells. Currently, FACS is limited to high complexity laboratories with time consuming pre-analytical steps, requiring trained technologists available only during business hours. The Accellix table top flow cytometer automates the process with results available within 20 minutes. Sample preparation and reading are performed in a dedicated disposable cartridge. Analytical data processing utilizing proprietary algorithms provides answers directly to the user.
Methods
The Accellix disposable cartridge-based platform implements sample preparations using three reagent blisters. The three Accellix CD64 cartridge blisters contain staining cocktail of conjugated monoclonal antibodies, lysis buffer and reference beads respectively. Once sample processing is complete, the sample flows through a dedicated reading channel where data is acquired. The present study compared the results of neutrophil CD64 levels performed by standard laboratory FACS and the Accellix system in ICU infected and control patients and normal volunteers.
Results
In a demonstration of cell surface marker quantitation a comparison study of 118 blood samples showed a correlation coefficient of 0.94 for Accellix determined neutrophil CD64 compared to those determined using a FACS. The comparison of the CD64 levels performed by the FACS and the Accellix system are shown in figure 1.
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Sprung, C., Morales, R., Kasdan, H. et al. Comparison of cd64 levels performed by the facs and accellix systems. ICMx 3 (Suppl 1), A1012 (2015). https://doi.org/10.1186/2197-425X-3-S1-A1012
Published:
DOI: https://doi.org/10.1186/2197-425X-3-S1-A1012