Accuracy of an automatic analysis software to detect microvascular density parameters

Introduction

Analysis of microvascular density parameters is time consuming and operator-dependent.1 This is the main limitation to use microvascular monitoring in clinical practice as a “point-of-care” tool. Recently, an automatic analysis software has been developed and could allow us to obtain results quickly.

Objectives

The aim of this study was to assess the accuracy of microvascular density parameters (total vessel density (TVD), perfused vessel density (PVD) and proportion of perfused vessels (PPV)) obtained by the new automatic analysis software CytoCamTools 1.7.12 (CC) (Braedius, Amsterdam, The Netherlands) in comparison with Automated Vascular Analysis (AVA) software 3.2 (MicroVision Medical, Amsterdam, The Netherlands).

Methods

Sublingual microcirculatory videos were obtained using an incidence dark field-imaging device (CytoCam, Braedius, Amsterdam, The Netherlands). Only videos with a high quality score2 were selected for the analysis. Each video was analysed using AVA 3.2 by two skilled operators and results were compared with the analysis obtained by the CytoCamTools 1.7.12 software. Bland-Altman analysis was used to look at the agreement between the automatic software and the operators.

Results

84 videos from 22 patients after cardiac surgery were analysed. The mean bias between TVD-CC and TVD-AVA was 2.17 mm/mm^2 (95% CI 1.44 to 2.90; p = 0.0001) with limits of agreement (LOA) of -4.41 (95% CI -5.66 to -3.16) and 8.76 (95% CI 7.50 to 10.01) mm/mm^2 (Figure 1). The mean bias between PVD-CC and PVD-AVA was 6.54 mm/mm^2 (95% CI 5.67 to 7.42; p < 0.0001) with LOA of -1.37 (95% CI -2.87 to 0.13) and 14.46 (95% CI 12.95 to 15.96) mm/mm^2 (Figure 2). A good correlation was found between the average of PPV-CC and PPV-AVA and the difference of the two measurements where ΔPPV = -1.85(mean PPV) +181 (R=-0.94, (95%CI -0.96 to -0.91; p < 0.0001).

Figure 1
figure 1

Bland-Altman between TVD-AVA and TVD-CC.

Figure 2
figure 2

Bland-Altman between PVD-AVA and PVD-CC.

Conclusions

This study shows acceptable bias but wide limits of agreement for the comparison of TVD, PVD and PPV between the automated CC system and skilled operators. Further software improvements may be needed before real time point of care testing of the microcirculation can be used at the bedside.

References

  1. De Backer D, Hollenberg S, Boerma C, Goedhart P, Büchele G, Ospina-Tascon G, et al: How to evaluate the microcirculation: report of a round table conference. Crit Care. 2007, 11 (5): R101-10.1186/cc6118.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Massey MJ, Larochelle E, Najarro G, Karmacharla A, Arnold R, Trzeciak S, et al: The microcirculation image quality score: development and preliminary evaluation of a proposed approach to grading quality of image acquisition for bedside videomicroscopy. J Crit Care. 2013, 28 (6): 913-917. 10.1016/j.jcrc.2013.06.015.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Carsetti, A., Pierantozzi, S., Aya, H. et al. Accuracy of an automatic analysis software to detect microvascular density parameters. ICMx 3 (Suppl 1), A415 (2015). https://doi.org/10.1186/2197-425X-3-S1-A415

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2197-425X-3-S1-A415

Keywords