Effects of Positive End Expiratory Pressure On Regional Ventilation-Perfusion Matching And Respiratory Mechanics: A Clinical Study

Introduction

In intubated acute respiratory failure patients, inhomogeneity of ventilation-perfusion matching (i.e., presence of over-ventilated under-perfused lung regions) might determine extreme pH shifts and increase the risk of developing ventilator-induced injury. Positive end-expiratory pressure (PEEP) redistributes tidal ventilation towards more perfused dependent lung regions, potentially improving homogeneity of ventilation-perfusion matching.

Objectives

In this study, we used Electrical Impedance Tomography (EIT) to assess global and regional lung ventilation and perfusion at different PEEP levels and to verify the effects of PEEP on homogeneity of ventilation-perfusion matching.

Methods

We enrolled 20 intubated critically ill patients undergoing controlled mechanical ventilation, sedated, paralyzed, with PaO2/FiO2≤300 mmHg and PEEP≥5 cmH2O. We randomly applied two PEEP levels (clinical and clinical + 5 cmH2O) for 20 minutes each and collected ventilation and EIT data at the end of each step. From EIT, we measured: 1. regional ventilation heterogeneity (VtHet, defined as the ratio between Vt reaching non-dependent/dependent lung); 2. regional homogeneity of ventilation-perfusion matching (HV/P); 3. regional compliance; 4. cumulated regional lung hyperdistension.

Results

Patients were 62 ± 12 years old, PaO2/FiO2 was 197 ± 52 mmHg, lower PEEP was 7 (7-9) cmH2O while higher PEEP was 12 (12-14) cmH2O (p < 0.001). At higher PEEP, VtHet was reduced (1.8 (1.5-2.4) vs. 2.2 (1.8-2.6), p < 0.001). Regional HV/P improved at higher PEEP in non-dependent areas (0.42 ± 0.24 vs. 0.29 ± 0.25, p < 0.01) as well as in the dependent ones, albeit non-significantly (0.37 ± 0.20 vs. 0.33 ± 0.24, p = 0.196) (Figures 1 and 2).

Figure 1
figure 1

H v/p non-dependent regions.

Figure 2
figure 2

H v/p dependent regions.

Finally, by applying higher PEEP, regional compliance of non-dependent lung decreased (31 ± 12 vs. 37 ± 13 mL/cmH2O, p < 0.001) and cumulative hyperdistension of the same areas significantly increased (+18 ± 7%, p < 0.001).

Conclusions

Improved homogeneity of ventilation-perfusion matching might represent one of the protective mechanisms associated with the use of higher PEEP. On the other hand, such benefits must be balanced with increased risk of hyperdistension of non-dependent lung.

Grant Aknowledgement

Institutional.

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access  This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Mauri, T., Eronia, N., Turrini et al. Effects of Positive End Expiratory Pressure On Regional Ventilation-Perfusion Matching And Respiratory Mechanics: A Clinical Study. ICMx 3 (Suppl 1), A8 (2015). https://doi.org/10.1186/2197-425X-3-S1-A8

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2197-425X-3-S1-A8

Keywords