Skip to main content
  • Research Articles
  • Open access
  • Published:

Bedside monitoring of lung volume available for gas exchange

Abstract

Background

Bedside measurement of lung volume may provide guidance in the personalised setting of respiratory support, especially in patients with the acute respiratory distress syndrome at risk of ventilator-induced lung injury. We propose here a novel operator-independent technique, enabled by a fibre optic oxygen sensor, to quantify the lung volume available for gas exchange. We hypothesised that the continuous measurement of arterial partial pressure of oxygen (PaO2) decline during a breath-holding manoeuvre could be used to estimate lung volume in a single-compartment physiological model of the respiratory system.

Methods

Thirteen pigs with a saline lavage lung injury model and six control pigs were studied under general anaesthesia during mechanical ventilation. Lung volumes were measured by simultaneous PaO2 rate of decline (VPaO2) and whole-lung computed tomography scan (VCT) during apnoea at different positive end-expiratory and end-inspiratory pressures.

Results

A total of 146 volume measurements was completed (range 134 to 1869 mL). A linear correlation between VCT and VPaO2 was found both in control (slope = 0.9, R2 = 0.88) and in saline-lavaged pigs (slope = 0.64, R2 = 0.70). The bias from Bland–Altman analysis for the agreement between the VCT and VPaO2 was − 84 mL (limits of agreement ± 301 mL) in control and + 2 mL (LoA ± 406 mL) in saline-lavaged pigs. The concordance for changes in lung volume, quantified with polar plot analysis, was − 4º (LoA ± 19°) in control and − 9° (LoA ± 33°) in saline-lavaged pigs.

Conclusion

Bedside measurement of PaO2 rate of decline during apnoea is a potential approach for estimation of lung volume changes associated with different levels of airway pressure.

Background

The acute respiratory distress syndrome (ARDS) is a common condition affecting about 100,000 patients yearly in the European Union and has a high mortality rate of ~ 40%, with ARDS patients occupying ~ 10% of intensive care unit beds in the UK [1]. ARDS patients require mechanical ventilation to maintain blood gases within the desired range. Despite low tidal volumes and limited plateau pressures being associated with better outcomes, this lifesaving mechanical ventilation can paradoxically contribute to the high mortality by further damaging the lung [2]. Clinical trials aimed at reducing lung injury via the application of high positive end-expiratory pressure (PEEP) levels have demonstrated conflicting results [3], suggesting that similar ventilator settings do not produce the same response at the alveolar level between different patients, especially in the presence of the “baby lung”, a smaller lung volume available for ventilation [4]. This could also be the case in the presence of lung heterogeneity like in COVID-19, where very similar arterial partial pressure of oxygen (PaO2) can be observed for very different computed tomography (CT) values [5,6,7]. There is a clear need for developments to enable individual titration of mechanical ventilation strategy, for example bedside real-time measurement of lung volume actually available for gas exchange, especially following changes between PEEP levels [8], which may contribute to the identification of patients with recruitable lungs.

Techniques currently available for the bedside measurement of lung volume are not routinely used because of different limitations including costs, physical dimensions, accuracy, time and conditions required for the measurement itself [9,10,11], although promising, operator-independent approaches are emerging [12,13,14,15,16].

In the presence of a single-compartment model of the lung, and for a known oxygen uptake, the rate of PaO2 decline during apnoea is inversely proportional to lung volume [17,18,19]. While the injured lung is highly heterogeneous in the dynamic state, this heterogeneity may be reduced after a few seconds of apnoea [20]. Continuous monitoring of PaO2 during breath-holding manoeuvres at the bedside may allow measurement of lung volume, and together with lung mechanics parameters, enable the calculation of lung strain ratio [21].

The aim of this study was to validate a novel method to determine lung volume available for gas exchange in real-time at the bedside by PaO2 rate of decline and oxygen uptake during apnoea. These measurements were performed at different positive end-expiratory and end-inspiratory pressures in mechanically ventilated pigs with saline lavage lung injury and in control pigs. Simultaneous whole-lung CT imaging data were used for comparison, where the lung volumes determined by CT would include regions with air trapped behind closed airways and aerated regions that are poorly perfused with blood, each of which would not contribute significantly to gas exchange, and could overlook regions with very limited aeration.

Methods

Ethical approval

The study was performed in a total of 19 domestic pigs (mean weight (SD) = 30 (2) kg) as part of different experiments at the Hedenstierna laboratory, Uppsala University, Sweden. The studies were approved by the regional animal welfare ethics committee (Ref: C98/16) and adhered to the Animal Research: Reporting of in vivo Experiments guidelines [22].

Animal preparation and monitoring

The animals’ baseline characteristics are summarised in Table 1. Animals received intramuscular sedation and were anaesthetised with total intravenous anaesthesia as described elsewhere [23]. Mechanical ventilation was delivered at 20–25 breaths per minute with a tidal volume (VT) of 10 mL/kg and an inspiratory:expiratory ratio (I:E) of 1:2 (Servo-I, Maquet Critical Care, Solna, Sweden). Lack of spontaneous movements, absence of reaction to painful stimulation between the front hooves and absence of cardiovascular signs of sympathetic stimulation (increases in heart rate or arterial blood pressure) confirmed the depth of anaesthesia. Once anaesthesia was ascertained, rocuronium was administered for muscle relaxation. Continuous infusion of Ringerfundin™ solution (Braun Melsungen Ag, Melsungen, Germany) was used for fluid replacement at 10 mL/kg/h during the preparation and at 7 mL/kg/h thereafter. Cardiac output was monitored intermittently by pulmonary artery catheter thermodilution, together with arterio-mixed venous blood samples used to calculate oxygen uptake \(\left({\dot{V}}_{{\mathrm{O}}_{2}}\right)\) via the Fick’s principle. Physiological parameters were continuously recorded with standard patient monitors (respiratory monitor: Datex Ohmeda Capnomac Ultima; multi-parameter patient monitor: Datex AS3) and electrical impedance tomography (EIT) to monitor pulmonary regional ventilation changes (Enlight, TIMPEL SA, São Paulo, Brazil). Analogue signals were continuously recorded on a computer via PowerLab (AD Instruments, New Zealand).

Table 1 Baseline characteristics for animals

Saline lavage lung injury model

For this initial proof-of-concept study, a collapse-prone lung injury was induced with a technique modified from Lachmann and colleagues [24]. Mechanical ventilation with a fraction of inspired O2 (FIO2) of 1.0 preceded the disconnection of the ventilator and surfactant depletion via lung lavage with 30 mL/kg of 0.9% saline solution (at 37 °C) instilled via the tracheal tube. After 30 s, saline was drained out of the lungs and ventilation recommenced. This process was repeated until the PaO2:FIO2 ratio (PFR) < 300 mmHg (40 kPa) was achieved at a PEEP of 5 cmH2O and FIO2 of 0.7.

Study protocol

A series of 20 s breath-holding manoeuvres was performed at different PEEP levels [range 0 to 20 cmH2O] and at end-inspiratory pressures associated with a VT of 10 mL/kg. Whole lung CT scans and continuous PaO2 data were recorded simultaneously during the breath-holding manoeuvres.

CT image acquisition and analysis

CT images were acquired with a Somatom Definition Flash (Siemens, Erlangen, Germany), used as a gold standard to measure lung volumes. Tube voltage was 80 kV, with 364 mA current and 64 × 60 mm collimation. Reconstituted voxel dimensions were 0.5 × 0.5 × 5 mm.

CT image analysis was performed using 3D Slicer v4.10.2 [25] (https://www.slicer.org), as presented elsewhere [26]. Aerated lung volume was calculated according to mean Hounsfield unit (HU) density within the segmented lung volume.

The HU boundaries were water with a density of 1 g cm−3 (0 HU) and air with a density of 0 g cm−3 (− 1000 HU). Gas volume was computed as in Chiumello et al. [27]:

$${V}_{\mathrm{CT}}= \frac{-\mathrm{CT} \,\mathrm{density} (\mathrm{HU})}{1000} \times \mathrm{segmented\, lung \,volume}.$$

Continuous measurement of PaO2 and mathematical modelling

PaO2 was recorded with a fibre optic sensor based on luminescence quenching by oxygen of a fluorophore embedded in a polymer material, polymethyl-methacrylate [23, 28,29,30,31], inserted in a carotid artery via a standard arterial catheter. The sensor has a response time ~ 100 ms and data were recorded continuously at 10 Hz. Technical properties of this sensor are presented elsewhere [30, 32].

As illustrated in Fig. 1, the lung was assumed to perform as a single compartment with a constant oxygen uptake. Here the rate of PaO2 decline is inversely proportional to lung volume, and directly proportional to the rate of oxygen uptake by the pulmonary circulation [17, 18]. Based on the equation presented in Fig. 1, predicted lung volume available for gas exchange can be calculated on the basis of measured oxygen uptake and rate of PaO2 decline during each breath-hold manoeuvre. The single compartment assumption was supported by stable airway pressure and flow, and EIT signal during the relevant period of apnoea, as illustrated in Fig. 2. In this single compartment assumption, the rate of alveolar oxygen change equals the difference between the rates of input via ventilation during inspiration and output via continuous uptake by the pulmonary circulation and elimination during expiration [19]. Pigs’ metabolism (e.g. heart rate, respiration and oxygen uptake) was stable during the experimental protocol.

Fig. 1
figure 1

Algorithm to calculate lung volume available for gas exchange. The text in the coloured boxes summarises the main steps in the algorithm leading to the equation used for the calculation of lung volume available for gas exchange. The text in the white boxes presents the physiological and signal conditions imposed a priori for the automatic processing and analysis of the raw PaO2 signal, necessary for a bias-free selection of the representative period that was used for the automatic calculation of lung volume available for gas exchange. Additional methodological details are presented in the methods. PaO2 arterial partial pressure of oxygen; SNR: signal-to-noise ratio; O2: oxygen uptake; Patm: atmospheric pressure; LV: lung volume

Fig. 2
figure 2

Representative responses to breath-holding manoeuvres performed during expiration and inspiration. a The arterial partial pressure of oxygen (PaO2), b the airway pressure, c the electrical impedance tomography (EIT) for three gravitational regions of interest (ROI), and d the airway flow. The vertical dashed red and black lines present the start and end points of the period used for analysis; minimum and maximum durations of this period were 2 and 5 s, depending on linearity of the PaO2 signal (see “Methods” for details)

The novel PaO2 signal processing technique proposed here is based on the calculation of PaO2 gradient between two random time points in the PaO2 signal during the overall breath-hold period, defined as the period with no airway flow. This iterative analysis was repeated a hundred times for each of the breath-hold manoeuvres, leading to the algorithm proposed here (Fig. 1), which identifies the period of linear decline on which the calculations to estimate lung volume available for gas exchange were based.

The linear PaO2 signal recorded in the last 2–5 s of each breath-holding manoeuvre was used to calculate the lung volume; an example is illustrated in Fig. 2. The airway flow, pressure and EIT signals were stable in the periods considered for analysis. The maximum analysis period was limited to 5 s to exclude transient changes in pulmonary ventilation, perfusion and their distributions, which might have occurred at the beginning of the breath-holding manoeuvres. These transient changes could have temporarily affected the linearity of the PaO2 rate of decline and would be overlooked by a single-compartment model of the respiratory system, hence would affect the accuracy of lung volume calculation. If the PaO2 decline was linear for 5 s, then a 5-s period was used to calculate the lung volume. Alternatively, a period shorter than 5 s was used for analysis, where a minimum period of at least 2 s was required because shorter periods were associated with relatively low signal-to-noise ratios and reduced accuracy and confidence. Breath-holding manoeuvres where PaO2 signal-to-noise ratio was < 30 dB were excluded from the analysis, as were those manoeuvres when PaO2 signal was smaller than 100 mmHg because of haemoglobin desaturation, which itself reduces the rate of PaO2 decline. For these two conditions combined, 20% of the data were excluded. Overall, this standardised procedure first filtered the raw signal to exclude cases where low signal-to-noise ratio and haemoglobin saturation reduction could have confounded the analysis, and then automatically calculated lung volume available for gas exchange, using an operator- and patient-independent approach to avoid subjectivity and bias.

Statistical analysis

Linear regression, Bland–Altman and polar plots were used to assess the relationship and agreement between absolute lung volumes and their changes measured with CT and with PaO2 rate of decline [33, 34]. The linear relationship between measurements was analysed using linear mixed effects modelling with variation caused by different animals considered as a random effect. Conditional R2 values based on the entire model are reported [35]. Change in lung volumes (ΔV) were calculated from their values measured at different airway pressure levels, and compared using four-quadrant and polar plots [36], where concordance was assumed adequate within ± 30º of the horizontal. Statistical analyses were performed in Matlab v2018b (Mathworks, MA, USA).

Results

A total number of 146 paired lung volume measurements were analysed in 6 control pigs and 13 saline lavage lung injury pig models at five different PEEP levels.

Figure 3 shows lung volumes and their PEEP-associated changes measured by PaO2 and CT. Overall, lung volumes available for gas exchange measured by PaO2 rate of decline (VPaO2) and those measured by whole-lung CT (VCT) increased with PEEP in the saline lavage model and in control animals. VPaO2 were greater than VCT at PEEP 0 cmH2O in the saline lavage lung injury model during end-expiratory (volume difference = 270(96) mL) and the associated end-inspiratory (volume difference = 319(234) mL) breath-holding manoeuvres. Additional file 1: Table S1 details these VPaO2 and VCT and their differences. Additional file 1: Table S2 shows details of multiple replicate measurements and that the mean coefficient of variation for VPaO2 repeated measurements within animal was less than 10% in both the control animals and in the saline lavage lung injury model.

Fig. 3
figure 3

Lung volume measurement (mean ± SD) from PaO2 analysis and CT imaging at different PEEP levels during incremental PEEP titration. VPaO2 and VCT measurements in the control animals during a apnoea at end-expiration and b end-inspiration. VPaO2 and VCT measurements in the saline lavage lung injury model during c apnoea at end-expiration and d end-inspiration. Additional file 1: Table S2 provides full details of volumes measured and probabilities of difference

Absolute lung volumes

Figure 4 shows the positive correlations between VPaO2 and VCT. Correlation values were 0.90 and 0.64, and R2 were 0.88 and 0.70, respectively, in the control pigs (Fig. 4a) and in the saline lavage lung injury model (Fig. 4c). The mean bias (± 95% limits of agreement) for VPaO2 and VCT was − 84 mL (± 301 mL) in control pigs (Fig. 4b), and 2 mL (± 405 mL) in the saline lavage lung injury model (Fig. 4d), where the mean bias tended to decrease at larger lung volumes.

Fig. 4
figure 4

Linear regression and Bland–Altman analyses for absolute volume measurements of VPaO2 and VCT in the control animals (a, b), and in the saline lavage lung injury model (c, d). Individual points represent a paired set of measurements at different PEEP levels (individual values in Additional file 1: Table S1). Each colour shows results from one animal. Circles and squares represent volumes measured, respectively, during end-expiratory and end-inspiratory apnoea. a, c Black solid lines are the regression lines, and the blue lines are the identity lines. b, d Black solid lines are the Bland–Altman plots’ regression lines, blue lines are the mean bias and red lines are the upper and lower limits of agreement (± 1.96 SD) with 95% confidence interval

Lung volume changes

Figure 5 shows 100% concordance and positive correlations between ΔVPaO2 and ΔVCT. Correlation values were 0.82 and 0.59, and R2 were 0.75 and 0.76, respectively, in the control pigs (Fig. 5a) and in the saline lavage lung injury model (Fig. 5c). The mean angular bias (± 95% radial limits of agreement) for polar agreement between VPaO2 and VCT was − 4° (± 19°), with a concordance of 97% in the control pigs (5b), and − 9° (± 33°), with a concordance of 86% in the saline lavage lung injury model (5d).

Fig. 5
figure 5

Volume changes (ΔVPaO2 and ΔVCT) in the control animals (a, b), and in the saline lavage lung injury model (c, d). Individual points represent a paired set of measurements and different colours represent changes associated with different ΔPEEP in each animal. a, c Black solid lines are the regression lines, and blue lines are the identity lines. b, d The stronger the agreement between two measurements, the closer the data are to the horizontal line (0°). b, d Blue lines are the mean angular bias, and red lines are the upper and lower limits of agreement; good concordance is assumed to be within ± 30° of the horizontal zero line. LoA limits of agreement

Discussion

This pre-clinical study presents a novel bedside method to monitor lung volume available for gas exchange, and it demonstrates the positive correlation, agreement and small bias between VPaO2 and VCT in a mechanically ventilated pig saline lavage lung injury model and in control pigs during breath-holding manoeuvres. While these results were associated with relatively small limits of agreement at low lung volumes, the limits of agreement indicated reduced accuracy at larger lung volumes, outside the < 5% recommendation [37]. The relatively smaller radial limits of agreement indicated a greater accuracy in the measurements of ΔVPaO2 and ΔVCT. VPaO2 results variability within animal was < 10%, within the reported acceptable limits of coefficient of variation [38].

The single-compartment model assumes that the air volume measured contributes to gas exchange in the lung (i.e. ventilated and perfused alveolar volume), whereas the air volume measured with CT imaging includes dead spaces. It is possible that this difference between techniques has contributed to the observed CT overestimation of absolute lung volume, especially at larger volumes in the saline lavage lung injury model, where greater volumes of air trapped behind closed airways may be present [39], as well as gas within conducting airways. At low lung volumes, especially at PEEP 0 cmH2O after saline lavage, it is possible that the HU range for non-aerated lung included a portion of lung that had at least a degree of ventilation and perfusion, hence contributed to gas exchange. Here, CT may not clearly distinguish between atelectatic regions not contributing to gas exchange, and regions with a very low ventilation and perfusion ratio (V/Q). For example, VCT at PEEP 5 cmH2O could include volumes from the ideal alveolus and regions with a high V/Q (i.e. alveolar dead space), while VPaO2 would include volumes from the ideal alveolus and regions with a low V/Q. When PEEP is reduced to 0 cmH2O, VCT would be reduced both by the loss of high V/Q regions, as well as by the greater proportion of atelectasis and low V/Q regions. In contrast, VPaO2 would still capture both ideal alveolus and also regions with low V/Q, possibly explaining the greater VPaO2 compared with VCT. This difference between techniques was smaller when considering changes in lung volume that may be important when determining, for example, the effect of a recruitment manoeuvre on lung volume.

The main limitations of the VPaO2 technique include the requirement for the brief ~20-s interruption of mechanical ventilation, the assumption that the lung can be mathematically modelled as a single compartment, for PaO2 to be greater than 100 mmHg, for an accurate measurement of oxygen uptake, and for a signal-to-noise ratio > 30 dB.

The VPaO2 technique must be performed during a breath-holding manoeuvre because rapid pulmonary ventilation and perfusion changes occur dynamically during tidal breathing, when the lung cannot be modelled as a single compartment. In this sense, a breath-holding manoeuvre shorter than ~ 20 s may be affected by transient and regional changes in pulmonary perfusion and ventilation, associated with pendelluft [40, 41]. Similarly long (or even longer) end-inspiratory breath-holding manoeuvres may be performed for recruitment in ARDS patients [42], but end-expiratory breath-holding manoeuvres, even at PEEP 5 cmH2O as in our study here, may increase the risk of greater lung collapse.

While PaO2 is normally greater than 100 mmHg in healthy, anaesthetised patients, it is typically lower in ARDS patients, where titration of mechanical ventilation is particularly important [2]. In order to reduce the chances of lung volume underestimation caused by haemoglobin desaturation, mean PaO2 would need to be raised above 100 mmHg for a few minutes, also as a safety precaution, for the duration of the breath-holding manoeuvre required for the measurement of lung volume via PaO2 rate of decline. This increase in PaO2 via greater FIO2 may not always be readily achievable.

Oxygen uptake was calculated from intermittent cardiac output measurements with thermodilution and from simultaneous arterio-venous blood samples [43]. In our study, a single oxygen uptake level was used to calculate VPaO2 across different breath-holding manoeuvres in order to limit the volume of saline injections and blood samples over the ~ 2-h experimental period. This approach could have contributed to the observed VPaO2 variability due to potentially overlooked changes in oxygen uptake, especially at higher lung volumes (end-inspiratory breath holds) in the saline lavage lung injury model. Here, greater positive airway pressure could have redistributed pulmonary perfusion [44] and reduced cardiac output during the breath-holding, hence reducing pulmonary oxygen uptake, possibly leading to an underestimation of lung volume. A greater accuracy would be expected with appropriate data collection timing, such as simultaneous measurement of oxygen uptake and VPaO2.

The fibre optic oxygen sensors used for our study are not yet produced in large quantities, hence their signal-to-noise ratio may be variable. Sensors’ positional changes within the arterial vessel could also reduce the signal-to-noise ratio, although these changes are unlikely to occur. Further developments in the manufacturing process of these prototype sensors are likely going to improve signal-to-noise ratio and generate more accurate VPaO2 measurements.

The main advantages of the VPaO2 technique include its potential use at the bedside, availability of lung volume measurement outcome within minutes, limited costs, and operator- and patient-independent results. The VPaO2 technique was more sensitive to relative than to absolute estimation of lung volume available for gas exchange. From a clinical perspective, this sensitivity may help distinguish recruitable from non-recruitable lung.

Conclusion

In conclusion, lung volumes estimated from the PaO2 rate of decline during breath-holding manoeuvres correlated with volumes measured by CT, but with large limits of agreement due to several confounders. The validity of this novel method needs to be confirmed in other lung injury models before it is used clinically.

Availability of supporting data and materials

Data are available upon reasonable request.

References

  1. Bellani G, Laffey JG, Pham T et al (2016) Epidemiology, patterns of care, and mortality for patients with acute respiratory distress syndrome in intensive care units in 50 countries. J Am Med Assoc 315:788–800. https://doi.org/10.1001/jama.2016.0291

    Article  CAS  Google Scholar 

  2. Slutsky AS, Ranieri VM (2013) Ventilator-induced lung injury. N Engl J Med 369:2126–2136

    Article  CAS  PubMed  Google Scholar 

  3. Amato MBP, Meade MO, Slutsky AS et al (2014) Driving pressure and survival in the acute respiratory distress syndrome. N Engl J Med 372:747–755. https://doi.org/10.1056/NEJMsa1410639

    Article  CAS  Google Scholar 

  4. Gattinoni L, Marini JJ, Pesenti A et al (2016) The “baby lung” became an adult. Intensive Care Med 42:663–673. https://doi.org/10.1007/s00134-015-4200-8

    Article  PubMed  Google Scholar 

  5. Gattinoni L, Chiumello D, Caironi P et al (2020) COVID-19 pneumonia: different respiratory treatments for different phenotypes? Intensive Care Med 46:1099–1102

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  6. Vasques F, Sanderson B, Formenti F et al (2020) Physiological dead space ventilation, disease severity and outcome in ventilated patients with hypoxaemic respiratory failure due to coronavirus disease 2019. Intensive Care Med. https://doi.org/10.1007/s00134-020-06197-x

    Article  PubMed  PubMed Central  Google Scholar 

  7. Vogel DJ, Formenti F, Retter AJ et al (2020) A left shift in oxyhaemoglobin dissociation curve in patients with severe COVID-19. Br J Haematol. https://doi.org/10.1111/bjh.17128

    Article  PubMed  PubMed Central  Google Scholar 

  8. Goligher EC, Kavanagh BP, Rubenfeld GD et al (2014) Oxygenation response to positive end-expiratory pressure predicts mortality in acute respiratory distress syndrome: a secondary analysis of the LOVS and express trials. Am J Respir Crit Care Med 190:70–76. https://doi.org/10.1164/rccm.201404-0688OC

    Article  PubMed  Google Scholar 

  9. Maisch S, Boehm SH, Weismann D et al (2007) Determination of functional residual capacity by oxygen washin–washout: a validation study. Intensive Care Med 33:912–916. https://doi.org/10.1007/s00134-007-0578-2

    Article  PubMed  Google Scholar 

  10. Di Marco F, Rota Sperti L, Milan B et al (2007) Measurement of functional residual capacity by helium dilution during partial support ventilation: In vitro accuracy and in vivo precision of the method. Intensive Care Med 33:2109–2115. https://doi.org/10.1007/s00134-007-0833-6

    Article  PubMed  Google Scholar 

  11. Heinze H, Eichler W (2009) Measurements of functional residual capacity during intensive care treatment: the technical aspects and its possible clinical applications. Acta Anaesthesiol Scand 53:1121–1130. https://doi.org/10.1111/j.1399-6576.2009.02076.x

    Article  CAS  PubMed  Google Scholar 

  12. Mountain JE, Santer P, O’Neill DP et al (2018) Potential for noninvasive assessment of lung inhomogeneity using highly precise, highly time-resolved measurements of gas exchange. J Appl Physiol 124:615–631. https://doi.org/10.1152/japplphysiol.00745.2017

    Article  CAS  PubMed  Google Scholar 

  13. Bikker IG, Leonhardt S, Bakker J, Gommers D (2009) Lung volume calculated from electrical impedance tomography in ICU patients at different PEEP levels. Intensive Care Med 35:1362–1367. https://doi.org/10.1007/s00134-009-1512-6

    Article  PubMed  PubMed Central  Google Scholar 

  14. Wolf GK, Arnold JH (2005) Noninvasive assessment of lung volume: respiratory inductance plethysmography and electrical impedance tomography. Crit Care Med 33:S163–S169. https://doi.org/10.1097/01.ccm.0000155917.39056.97

    Article  PubMed  Google Scholar 

  15. Crockett DC, Tran MC, Formenti F et al (2020) Validating the inspired sinewave technique to measure the volume of the ‘baby lung’ in a porcine lung-injury model. Br J Anaesth 124:345–353. https://doi.org/10.1016/j.bja.2019.11.030

    Article  CAS  PubMed  Google Scholar 

  16. Tran MC, Crockett DC, Formenti F et al (2020) Lung heterogeneity and deadspace volume in acute respiratory distress syndrome animals using the inspired sinewave test. Physiol Meas. https://doi.org/10.1088/1361-6579/abc0b5

    Article  PubMed  Google Scholar 

  17. Dubois AB (1952) Alveolar CO2 and O2 during breath holding, expiration, and inspiration. J Appl Physiol 5:1–12. https://doi.org/10.1152/jappl.1952.5.1.1

    Article  CAS  PubMed  Google Scholar 

  18. Dubois AB, Britt AG, Fenn WO (1952) Alveolar CO2 during the respiratory cycle. J Appl Physiol 4:535–548. https://doi.org/10.1152/jappl.1952.4.7.535

    Article  CAS  PubMed  Google Scholar 

  19. Formenti F, Bommakanti N, Chen R et al (2017) Respiratory oscillations in alveolar oxygen tension measured in arterial blood. Sci Rep 7:1–10. https://doi.org/10.1038/s41598-017-06975-6

    Article  CAS  Google Scholar 

  20. Markstaller K, Eberle B, Kauczor HU et al (2001) Temporal dynamics of lung aeration determined by dynamic CT in a porcine model of ARDS. Br J Anaesth 87:459–468. https://doi.org/10.1093/bja/87.3.459

    Article  CAS  PubMed  Google Scholar 

  21. Cortes GA, Marini JJ (2013) Two steps forward in bedside monitoring of lung mechanics: transpulmonary pressure and lung volume. Crit Care 17:219. https://doi.org/10.1186/cc12528

    Article  PubMed  PubMed Central  Google Scholar 

  22. Kilkenny C, Browne WJ, Cuthill IC et al (2013) Improving bioscience research reporting: the arrive guidelines for reporting animal research. Animals 4:35–44. https://doi.org/10.3390/ani4010035

    Article  Google Scholar 

  23. Crockett DC, Cronin JN, Bommakanti N et al (2019) Tidal changes in PaO2 and their relationship to cyclical lung recruitment/derecruitment in a porcine lung injury model. Br J Anaesth 122:277–285. https://doi.org/10.1016/j.bja.2018.09.011

    Article  CAS  PubMed  Google Scholar 

  24. Lachmann B, Robertson B, Vogel J (1980) In vivo lung lavage as an experimental model of the respiratory distress syndrome. Acta Anaesthesiol Scand 24:231–236. https://doi.org/10.1111/j.1399-6576.1980.tb01541.x

    Article  CAS  PubMed  Google Scholar 

  25. Fedorov A, Beichel R, Kalpathy-Cramer J et al (2012) 3D Slicer as an image computing platform for the Quantitative Imaging Network. Magn Reson Imaging 30:1323–1341. https://doi.org/10.1016/j.mri.2012.05.001

    Article  PubMed  PubMed Central  Google Scholar 

  26. Cronin JN, Borges JB, Crockett DC et al (2019) Dynamic single-slice CT estimates whole-lung dual-energy CT variables in pigs with and without experimental lung injury. Intensive Care Med Exp 7:59. https://doi.org/10.1186/s40635-019-0273-y

    Article  PubMed  PubMed Central  Google Scholar 

  27. Chiumello D, Cressoni M, Chierichetti M et al (2008) Nitrogen washout/washin, helium dilution and computed tomography in the assessment of end expiratory lung volume. Crit Care 12:2–9. https://doi.org/10.1186/cc7139

    Article  Google Scholar 

  28. Formenti F, Chen R, McPeak H et al (2014) A fibre optic oxygen sensor that detects rapid PO2 changes under simulated conditions of cyclical atelectasis in vitro. Respir Physiol Neurobiol 191:1–8. https://doi.org/10.1016/j.resp.2013.10.006

    Article  PubMed  PubMed Central  Google Scholar 

  29. Formenti F, Chen R, McPeak H et al (2015) Intra-breath arterial oxygen oscillations detected by a fast oxygen sensor in an animal model of acute respiratory distress syndrome. Br J Anaesth 114:683–688. https://doi.org/10.1093/bja/aeu407

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  30. Chen R, Formenti F, McPeak H et al (2016) Experimental investigation of the effect of polymer matrices on polymer fibre optic oxygen sensors and their time response characteristics using a vacuum testing chamber and a liquid flow apparatus. Sens Actuators B Chem 222:531–535. https://doi.org/10.1016/j.snb.2015.08.095

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  31. Chen R, McPeak H, Formenti F, et al (2013) Optimizing sensor design for polymer fibre optic oxygen sensors. In: Proceedings of IEEE Sensors. pp 1–4

  32. Chen R, Formenti F, McPeak H et al (2014) Optimizing design for polymer fiber optic oxygen sensors. Sensors J IEEE 14:3358–3364

    Article  CAS  Google Scholar 

  33. Nawarathna LS, Choudhary PK (2013) Measuring agreement in method comparison studies with heteroscedastic measurements. Stat Med 32:5156–5171. https://doi.org/10.1002/sim.5955

    Article  PubMed  Google Scholar 

  34. Bland JM, Altman DG (2007) Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat 17:571–582. https://doi.org/10.1080/10543400701329422

    Article  PubMed  Google Scholar 

  35. Nakagawa S, Schielzeth H (2013) A general and simple method for obtaining R2 from generalized linear mixed-effects models. Methods Ecol Evol 4:133–142. https://doi.org/10.1111/j.2041-210x.2012.00261.x

    Article  Google Scholar 

  36. Critchley LA, Lee A, Ho AMH (2010) A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg 111:1180–1192. https://doi.org/10.1213/ANE.0b013e3181f08a5b

    Article  PubMed  Google Scholar 

  37. Robinson PD, Latzin P, Verbanck S et al (2013) Consensus statement for inert gas washout measurement using multiple- and single breath tests. Eur Respir J 41:507–522. https://doi.org/10.1183/09031936.00069712

    Article  CAS  PubMed  Google Scholar 

  38. Wanger J, Clausen JL, Coates A et al (2005) Standardisation of the measurement of lung volumes. Eur Respir J 26:511–522. https://doi.org/10.1183/09031936.05.00035005

    Article  CAS  PubMed  Google Scholar 

  39. Derosa S, Borges JB, Segelsjö M et al (2013) Reabsorption atelectasis in a porcine model of ARDS: regional and temporal effects of airway closure, oxygen, and distending pressure. J Appl Physiol 115:1464–1473. https://doi.org/10.1152/japplphysiol.00763.2013

    Article  CAS  PubMed  Google Scholar 

  40. Otis AB, McKerrow CB, Bartlett RA et al (1956) Mechanical factors in distribution of pulmonary ventilation. J Appl Physiol 8:427–443. https://doi.org/10.1152/jappl.1956.8.4.427

    Article  CAS  PubMed  Google Scholar 

  41. Tabuchi A, Nickles HT, Kim M et al (2016) Acute lung injury causes asynchronous alveolar ventilation that can be corrected by individual sighs. Am J Respir Crit Care Med 193:396–406. https://doi.org/10.1164/rccm.201505-0901OC

    Article  PubMed  Google Scholar 

  42. Meade MO, Cook DJ, Guyatt GH et al (2008) Ventilation strategy using low tidal volumes, recruitment maneuvers, and high positive end-expiratory pressure for acute lung injury and acute respiratory distress syndrome: a randomized controlled trial. J Am Med Assoc 299:637–645. https://doi.org/10.1001/jama.299.6.637

    Article  CAS  Google Scholar 

  43. Stock MC, Ryan ME (1996) Oxygen consumption calculated from the Fick equation has limited utility. Crit Care Med 24:86–90. https://doi.org/10.1097/00003246-199601000-00015

    Article  CAS  PubMed  Google Scholar 

  44. Cronin JN, Crockett DC, Farmery AD et al (2020) Mechanical ventilation redistributes blood to poorly ventilated areas in experimental lung injury. Crit Care Med 48:e200–e208. https://doi.org/10.1097/CCM.0000000000004141

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

We are grateful to Oxford Optronix and to the staff at the Hedenstierna Laboratoriet and Radiology Department, Uppsala University Hospital, including Agneta Roneus, Kerstin Ahlgren, Mariette Anderson, Liselotte Pihl, Maria Swälas and Monica Segelsjö for their expertise and technical assistance.

Funding

FF was supported by the Medical Research Council (MC_PC_17164) and The Physiological Society (Formenti 2018), AL and GH were supported by the Swedish Lung and Heart Foundation (20170531) and the Swedish Research Council (K2015-99X-2273101-4).

Author information

Authors and Affiliations

Authors

Contributions

DCC, JNC and FF designed the experiments. DCC, JNC, JBB, MCT and FF carried out the experiments. MCT, DCC and JNC analysed the data. MCT, DCC, JNC, GH, AL, ADF and FF interpreted the data. AL, GH and FF contributed to financial support. MCT and FF wrote the manuscript. MCT, DCC, JNC, JBB, GH, AL, ADF and FF critically revised the manuscript. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Minh C. Tran or Federico Formenti.

Ethics declarations

Ethics approval and consent to participate

The studies were approved by the regional animal welfare ethics committee (Ref: C98/16) and adhered to the Animal Research: Reporting of in vivo Experiments guidelines.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Table S1.

Lung volume (upper part) and its changes (lower part) in both the control animals and the saline lavage lung injury model. Table S2. Mean (standard deviation) of the replicated volume measurements in 9 animals. An additional 50 measurements were performed, but data were not analysed due to either PaO2 < 100 mmHg, or signal-to-noise ratio < 30 dB. n: number of observations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Tran, M.C., Crockett, D.C., Cronin, J.N. et al. Bedside monitoring of lung volume available for gas exchange. ICMx 9, 3 (2021). https://doi.org/10.1186/s40635-020-00364-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40635-020-00364-6

Keywords