The ADVanced Organ Support (ADVOS) hemodialysis system removes IL-6: an in vitro proof-of-concept study

Abstract

Background

IL-6 is a pleiotropic cytokine modulating inflammation and metabolic pathways. Its proinflammatory effect plays a significant role in organ failure pathogenesis, commonly elevated in systemic inflammatory conditions. Extracorporeal blood purification devices, such as the Advanced Organ Support (ADVOS) multi hemodialysis system, might offer potential in mitigating IL-6's detrimental effects, yet its efficacy remains unreported.

Methods

We conducted a proof-of-concept in vitro study to assess the ADVOS multi system's efficacy in eliminating IL-6. Varying concentrations of IL-6 were introduced into a swine blood model and treated with ADVOS multi for up to 12 h, employing different blood and concentrate flow rates. IL-6 reduction rate, clearance, and dynamics in blood and dialysate were analyzed.

Results

IL-6 clearance rates of 0.70 L/h and 0.42 L/h were observed in 4 and 12-h experiments, respectively. No significant differences were noted across different initial concentrations. Reduction rates ranged between 40 and 46% within the first 4 h, increasing up to 72% over 12 h, with minimal impact from flow rate variations. Our findings suggest that an IL-6-albumin interaction and convective filtration are implicated in in vitro IL-6 elimination with ADVOS multi.

Conclusions

This study demonstrates for the first time an efficient and continuous in vitro removal of IL-6 by ADVOS multi at low blood flow rates. Initial concentration-dependent removal transitions to more consistent elimination over time. Further clinical investigations are imperative for comprehensive data acquisition.

Background

Critical care medicine has witnessed remarkable advancements in recent years, yet the management of critically ill patients with organ failures remains a formidable challenge. Among the complex web of inflammatory mediators orchestrating the systemic inflammatory response in these patients, interleukin-6 (IL-6) has garnered significant attention for its potential role in exacerbating organ dysfunction and adversely affecting clinical outcomes. In this context, extracorporeal blood purification devices, have emerged as innovative tools with the potential to mitigate the detrimental effects of excessive IL-6 [1].

IL-6 is a pleiotropic cytokine affecting inflammation and metabolic pathways. While IL-6 can exert positive effects, such as promoting an anti-inflammatory state of macrophages, limiting atheroma formation, or mediating insulin-sensitizing effects of physical exercise [2], its recent therapeutic interest is based on its negative effects. In its proinflammatory role, IL-6 is known to contribute to the pathogenesis of organ failure, coagulopathy, and immune dysregulation in critically ill patients and has been shown to be elevated in various acute systemic inflammatory syndromes and secondary organ dysfunctions (i.e., renal, hepatic or pulmonary) [3]. Increased levels of IL-6 induce the expression of various genes involved in inflammation, cell survival, and differentiation, which might result in the promotion of fever, acute phase response, endothelial activation, coagulation dysfunction, tissue injury, and organ dysfunction [3]. This makes IL-6 an attractive target for therapeutic intervention.

In this regard, different therapeutic strategies may be used, such as corticosteroids, antiviral agents, anti-cytokine antibodies or inhibitors (e.g., tocilizumab, anakinra, sarilumab), immunomodulators (e.g., baricitinib, ruxolitinib), convalescent plasma therapy, and extracorporeal blood purification (EBP) [4]. However, a significant knowledge gap exists concerning the direct elimination of IL-6 from the bloodstream using EBP. While several cytokine adsorption devices have reported IL-6 removal, the clinical efficacy of these interventions is being currently debated [5, 6]. Moreover, very few studies exist on the reduction of IL-6 levels using EBP systems that do not incorporate adsorptive materials or devices.

Among the plethora of therapies for critically ill patients, the ADVOS multi hemodialysis system has emerged as a promising medical device for the support of the liver, the kidney and the lung [7]. Based on the principle of albumin dialysis and using a customizable recirculating dialysate, ADVOS has reported the removal of protein-bound and water-soluble toxins while allowing a fluid-based CO2 removal and the correction of acid–base balance [8,9,10,11,12,13,14,15]. Understanding the potential of this device to remove IL-6 could represent a groundbreaking advancement in the field of critical care medicine, offering a novel approach to ameliorate the systemic inflammatory response associated with organ failure.

The primary aim of this study is to bridge the existing knowledge gap by investigating the efficacy of the ADVOS multi system in eliminating IL-6 from human blood in an in vitro setting. Our research seeks to determine the IL-6 removal kinetics of the ADVOS multi system. For this purpose, several scenarios using various IL-6 concentrations were tested. By comprehensively addressing these questions, we aim to contribute to a deeper understanding of the potential role of the ADVOS therapy in targeting IL-6, ultimately paving the way for novel therapeutic strategies that may enhance the management and outcomes of critically ill patients with organ failure.

Methods

Blood model

Fresh porcine blood was sourced from a local slaughterhouse (Münchner Schlachthof Betriebs GmbH, Munich, Germany) and processed following a standardized protocol. The blood was mixed with a modified Ringer's solution to attain a hematocrit level of 36%, maintaining standard electrolyte concentrations and typical blood gas values. To prevent coagulation, we administered 30,000 IU of heparin per liter of blood (Ratiopharm, Ulm, Germany). The blood was maintained at a constant temperature of 37 °C and gently agitated at a rate of 130 rpm to ensure its suitability for our intensive care research.

Different concentrations of IL-6 were spiked in blood. Briefly, recombinant human IL-6 (PeproTech, Cranbury, NJ, USA) was prepared according to the manufacturer's instructions for handling and reconstitution. The IL-6 was initially reconstituted in distilled water by gently shaking the vial after centrifugation. The IL-6 reconstituted solution was stored at 2 °C for short term use (up to 1 week) or at − 20 °C to ensure long term stability. During experiments, 0.5, 2.5, 10 and 50 µg of IL-6 were spiked into 5 L blood to achieve IL-6 concentrations of 100, 500, 2000 and 10,000 pg/mL, respectively.

ADVOS multi-hemodialysis system

The ADVOS multi (ADVITOS GmbH, Munich, Germany) is a hemodialysis system intended for the removal of water-soluble and protein-bound substances, for the correction of blood composition in case of electrolyte imbalance, including metabolic or hypercapnic acidosis, and for the removal of fluid, if needed. It consists of three interconnected circuits (Fig. 1). Briefly, bloods flows (100–400 mL/min) through the extracorporeal circuit, which bears two ELISIO 19H dialyzers (Nipro D.Med Germany GmbH, Hamburg, Germany). Here, human albumin enriched (200 mL, 20%) dialysate flowing from the dialysate circuit (800 mL/min) receives toxins from blood. The toxin-loaded dialysate fluid enters and recirculates then through the ADVOS multi regeneration circuit. Here, by applying pH and temperature changes, dialysate albumin adapts its structure to release the protein-bound toxins. These are then filtered by convection via two ELISIO 13H filters (Nipro D. Med Germany GmbH, Hamburg, Germany) into the waste together with the water-soluble toxins. The removed volume is replenished with fresh dialysate (160–320 mL/min), which is obtained by continuously mixing osmosis water, an alkaline (i.e., mainly NaOH), and an acidic concentrate (i.e., mainly HCl) online. Finally, the acidic to alkaline concentrate ratio defines the individualized dialysate pH (7.2–9.5).

Fig. 1
figure 1

Schematic representation of ADVOS multi and the experimental setting. A blood pool of 5 L was spiked with different concentrations of IL-6 and was subsequently treated with ADVOS multi. The ADVOS hemodialysis system consists of an extracorporeal circuit, a dialysate circuit and an albumin regeneration circuit (i.e., ADVOS multi circuit)

Additionally, in order to maintain adequate and physiological blood gas values, CO2 was continuously infused into the blood pool via an additional ELISIO 19H dialyzer (Nipro D.Med Germany GmbH, Hamburg, Germany) connected to a CO2 gas supply (Linde AG, Munich, Germany), as previously described [13].

Experimental set-up

In a preliminary analysis to rule out a spontaneous change of IL-6 levels, blood was spiked with a known concentration of IL-6. Blood was continuously stirred for 4 h without further treatment. Duplicate samples were taken and analyzed at different timepoints. Blood without IL-6 addition was used as negative control.

Once stated that no decomposition of IL-6 occurred in our blood model, blood was subjected to 4-h treatments with ADVOS multi. Table 1 shows the different settings that were employed with the aim to determine the effect of varying IL-6 concentrations, and blood and concentrate flows on the reduction rate and clearance of IL-6. Finally, to state if IL-6 could be continuously removed beyond 4 h, treatments lasting 12 h were conducted with a starting concentration of 10,000 pg/mL IL-6.

Table 1 Experimental settings during ADVOS multi treatments

Blood samples were taken in duplicate at minutes 0, 15, 30, 60, 120, 180, and 240 (4 h), and additionally at minutes 480, 720 (12 h) for the longer experiments. Each experiment was performed 3 times.

IL-6 measurement and removal analysis

Blood and dialysate samples (5 mL) were obtained at the specified time points. Blood was centrifuged at 4000 rpm for 15 min at 4 °C and 1 mL of the blood plasma was stored at − 25 °C until analysis. Dialysate did not undergo centrifugation. The Elecsys IL-6 sandwich chemiluminescent immunoassay (Roche Diagnostics GmbH, Mannheim, Germany) was used for sample analysis. Briefly, a sample of 18 µL is initially mixed with IL-6 specific antibodies and then incubated with ruthenium labelled IL-6 specific antibodies to form a sandwich complex. The complexes are then magnetically captured, inducing a chemiluminescent emission proportional to IL-6 concentration.

Reduction rate and clearance

Reduction rate was calculated as stated in Eq. 1, where Cf refers to the final IL-6 concentration (e.g., at 240 or 720 min for 4 and 12 h experiments, respectively), and C0 is the started spiked IL-6 concentration.

$$\text{Reduction rate} \left(\%\right)=1-\frac{{C}_{f}}{{C}_{0}} \times 100$$
(1)

Equation 2 was employed for the calculation of the elimination constant (Ke), where tf and t0 reflect the final (i.e., 4 or 12 h) and the starting (i.e., 0 min) timepoints, respectively. Ke was calculated assuming a first order kinetics.

$${K}_{e}({h}^{-1})=\frac{\text{ln}(\frac{{C}_{0}}{{C}_{f}})}{({t}_{f}-{t}_{0})}=\frac{\text{ln}{C}_{0}-\text{ln}{C}_{f}}{({t}_{2}-{t}_{1})}$$
(2)

For clearance (CL) calculation, a volume of distribution (Vd) of 5 L (i.e., blood volume) was set within Eq. 3.

$$\text{CL} (l/h)=\text{Vd}\times {K}_{e}$$
(3)

Statistical analysis

Data were documented and analyzed using Microsoft Excel and IBM SPSS 28.0 for Windows®, respectively. Reduction rate, and clearance are presented as mean and standard deviation (SD). An analysis of variance (ANOVA) followed by Bonferroni tests was conducted to determine differences between varying treatment settings. A p value lower than 0.05 was considered to indicate statistical significance.

Results

In the preliminary test, IL-6 was shown to be stable in the blood model with a statistically non-significant mean level change of 5% (data not shown). No IL-6 was detected in the negative control.

During treatments with ADVOS multi, a continuous removal of IL-6 was demonstrated. For all the settings tested, the course of IL-6 in blood experienced a fast decrease during the first 30 min and then continued to be removed in a more stable manner for the whole observational period (Fig. 2). Concomitantly, dialysate IL-6 levels increased first to then correlate with the decrease of IL-6 observed in blood (Fig. 2).

Fig. 2
figure 2

Course of IL-6 in blood (blue lines) and dialysate (orange lines) during 4 h treatment with ADVOS multi at varying starting concentrations. A 100 pg/mL; B 500 pg/mL; C 2,000 pg/mL; D 10,000 pg/mL. Error bars represent SD

The mean clearance of IL-6 with ADVOS multi varied between 0.42 L/h and 0.70 L/h (Fig. 3, orange columns). No significant differences between 4 h experiments at varying starting concentrations (p = 0.444) were detected. A statistical difference between different flow combinations was only detected in 12 h experiments (0.42 vs. 0.53 L/h, p = 0.003), but not in 4 h experiments (0.62 vs 0.70 l/h, p = 0.116) with blood flow 100 mL/min combined with concentrate flow 160 mL/min, and with blood flow 300 mL/min combined with concentrate flow 320 mL/min, respectively.

Fig. 3
figure 3

IL-6 reduction rate (blue columns) and clearance (orange columns) mean values for the different settings tested during ADVOS multi treatments. Error bars represent SD. Significance levels obtained by ANOVA are shown for each pair/group

The calculated clearance resulted in a mean reduction rate of IL-6 with ADVOS multi between 40 and 46% for the first 4 h and up to 72% for 12 h treatments (Fig. 3, blue columns). No significant statistical differences were observed among the varying concentrations tested (p = 0.274). The increase of blood flow and concentrate flow did not significantly change the removal rate of IL-6 in the 4 h treatment (43% vs. 43%, p = 0.454) but was statistically significant in 12 h treatments (67% vs. 72%, p = 0.016).

Discussion

Key results

This proof-of-concept work shows for the first time that the treatment with ADVOS multi can efficiently and continuously remove IL-6 from blood in an in vitro setting. ADVOS multi was able to remove up to 72% of the spiked IL-6 during 12 h of treatment. Moreover, removal rate was comparable for all the starting concentrations tested (100–10,000 pg/mL) and increasing blood and concentrate flows barely improved IL-6 elimination.

Interpretation

The ADVOS hemodialysis system is intended for patients with multiple organ failure suffering from acute, chronic and acute-on-chronic liver failure and/or renal failure and/or acidosis [7]. In these patients, especially in the presence of acidosis [16], a hyperinflammatory state with elevated levels of IL-6 might be present [17,18,19,20]. The mitigation of the hyperinflammation through the removal of IL-6 has been suggested as a therapeutic option [21].

IL-6 is a middle-molecule of about 24.5 kDa. Due to its size, it is not expected to be removed through conventional high-flux dialyzers [22]. ADVOS multi uses two polyethersulfone high flux hemodialyzers accounting for a surface of 3.8 m2 (i.e., 2 × 1.9 m2) in the extracorporeal circuit. In a hemodialysis setting, these filters have shown a good removal rate for small molecules [23], but a limited removal of large to medium molecules is expected [24, 25], especially at molecular weights above 20 kDa [23]. Even if limited through its size, one could expect that at higher initial IL-6 concentrations, higher gradients and consequently higher reductions rates could be achieved, attending to Fick’s Law [26]. This would be true for a diffusion-driven-elimination. However, the degree of elimination is not solely determined by the cytokine's molecular weight; instead, it's influenced by factors like charge, hydrophilicity, or carrier binding [27]. As demonstrated by other authors, cytokine removal is not only a diffusive process, and convection, and/or adsorption need to be considered [28]. In our experiments, we did not use any adsorptive membranes and changes in blood flow or initial concentrations did not exert any influence on the IL-6 reduction rate with ADVOS (Fig. 3). This points towards additional elimination mechanisms.

In this regard, an albumin-mediated IL-6 removal has been hypothesized based on a demonstrated in vitro interaction between albumin and IL-6 [29]. The authors of the study suggest that the presence of albumin within the dialysate facilitates an increased passage of cytokines over membrane pores and increases the affinity for albumin-associated cytokines on the dialysate side.

On top of a possible impact of dialysate albumin, convective transport might also play a role. ADVOS multi is a hemodialysis system and diffusive solute removal is, therefore, expected in the extracorporeal circuit. In addition to this, in the dialysate regeneration circuit, convective filtration occurs through two high flux filters of 1.3 m2 each. Concentrate flow refers to the amount of dialysate that is filtered and replaced. In detail, the amount of new dialysate entering the circuit accounts for the sum of volumes/flows of fresh permeate supply, of new BASE concentrate supply and of new ACID concentrate supply. This volume of fresh dialysate is compensated with the same volume of dialysate exiting the circuit into the waste at the same flow (i.e., concentrate flow). In our work, concentrate flows of 160 and 320 mL/min were tested, which could have been responsible for convective removal. Glancey et al. reported this phenomenon as increasing values of ultrafiltration during hemodiafiltration with high flux filters resulted in higher removal rates for low molecular weight proteins, including ß2-Microglobulin, Myoglobin and IL-6 [30].

The role of the dialysate regeneration circuit in ADVOS multi has additional particularities. As described above, the toxin-loaded dialysate enters the regeneration circuit where dialysate albumin unloading occurs and both protein-bound and water-soluble toxins are filtered by convection in exchange for fresh dialysate. However, a small proportion of dialysate does not enter this circuit, which allows to maintain a significant level of dialysate constituents. This works as a safety mechanism to avoid an excessive loss of relevant substances. Nevertheless, it allows a sufficient concentration gradient for an efficient continuous removal, as shown in Fig. 2. Although this might be seen as a disadvantage in comparison to conventional single-pass systems, it has been shown to be important to avoid the loss of anti-inflammatory cytokines [31].

Indeed, this work by Kaps et al. was the first to report data on cytokine removal with ADVOS multi [31]. The authors analyzed a panel of pro- and anti-inflammatory cytokines before and after a single ADVOS treatment in patients with acute-on-chronic liver failure (ACLF) and concluded that concentrations of pathomechanistically relevant cytokines remained unchanged. On one hand, the authors did not analyze levels before and after the dialyzers. Therefore, removal of interleukins cannot be ruled out, even if systemic plasma levels did not significantly change before and after treatment. An absence of correlation between IL-6 removal rate and reduction of systemic levels has already been documented for other blood purification devices [32]. On the other hand, the recirculation occurring within the ADVOS multi device could hinder an excessive removal of cytokines, when levels are close to physiologic values, which can be as high as 43.5 pg/mL for IL-6, as documented in a recent meta-analysis [33].

Generalizability

Attempts to reduce the systemic levels of IL-6 by other blood purification systems in a clinical setting has largely been documented [6, 34]. However, the focus on middle to large molecule removal is based either on adsorptive devices with the highest reduction rates in vitro [28], or membranes with high- and medium-cut-off (HCO and MCO, respectively) [23,24,25, 35,36,37]. When utilizing HCO or MCO membranes for convective therapy, it's essential to prevent significant albumin leakage during a renal replacement therapy session. While increasing the convection volume enhances middle molecule removal, a larger volume also carries a higher risk of albumin leakage [38].

Our work shows that ADVOS multi can remove IL-6 at low blood flows using conventional high flux filters, which minimizes albumin loss. Being this said, the main goal of the ADVOS therapy is not the sole removal of IL-6, for which other devices such as HCO or adsorptive membranes have been specifically developed. The ADVOS hemodialysis system seeks to restore the homeostasis of the patient through a multiorgan approach by efficiently removing water-soluble and protein-bound toxins, and by improving the blood composition through acidosis correction and CO2 removal [8,9,10,11,12,13]. Therefore, clinical data are needed to demonstrate the impact of IL-6 removal with ADVOS multi in the outcome of critically ill patients with multiple organ failure. Furthermore, the consequences that IL-6 removal and a possible concomitant depletion of molecules of similar characteristics through extracorporeal blood purification is still a matter of debate [39]. Thus, we suggest monitoring the clinical course of IL-6 in patients treated with ADVOS and the inclusion of the routine measurement of this cytokine as an endpoint in future clinical trials.

Limitations

Our study is limited by it is an in vitro nature and its applicability to the clinical setting might be restricted. Additionally, the results are based on short experiments that do not reflect the expected duration of a treatment. However, we have used a well-established blood model that allows to establish proof-of-concept data for the removal of relevant toxins, as shown previously [13, 40]. Furthermore, the design of the experiments with multiple sample extraction during the first hours allows a reliable estimation of the expected removal of IL-6.

Conclusion

In summary, our study demonstrates the efficient and continuous removal of IL-6 by ADVOS multi, even at low blood flow rates, as evidenced in vitro. We observed concentration-dependent removal within the initial timepoints, followed by a more consistent elimination in the subsequent hours. Recirculation of the dialysate proved effective in preventing the removal of low IL-6 concentrations. However, further investigation, both in vitro and in clinical settings, is essential to gather comprehensive and meaningful data.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

ACLF:

Acute-on-chronic liver failure

ADVOS:

Advanced organ support system

EBP:

Extracorporeal blood purification

HCO:

High-cut-off

IL-6:

Interleukin-6

MCO:

Medium-cut-off

References

  1. Ronco C, Bellomo R (2022) Hemoperfusion: technical aspects and state of the art. Crit Care 26:135. https://doi.org/10.1186/s13054-022-04009-w

    Article  PubMed  PubMed Central  Google Scholar 

  2. Fuster JJ, Walsh K (2014) The good, the bad, and the ugly of interleukin-6 signaling. EMBO J 33:1425–1427. https://doi.org/10.15252/embj.201488856

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  3. Fajgenbaum DC, June CH (2020) Cytokine storm. N Engl J Med 383:2255–2273. https://doi.org/10.1056/NEJMra2026131

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  4. Bhaskar S, Sinha A, Banach M et al (2020) Cytokine storm in COVID-19—immunopathological mechanisms, clinical considerations, and therapeutic approaches: The REPROGRAM Consortium Position Paper. Front Immunol. https://doi.org/10.3389/fimmu.2020.01648

    Article  PubMed  PubMed Central  Google Scholar 

  5. Becker S, Lang H, Vollmer Barbosa C et al (2023) Efficacy of CytoSorb®: a systematic review and meta-analysis. Crit Care 27:215. https://doi.org/10.1186/s13054-023-04492-9

    Article  PubMed  PubMed Central  Google Scholar 

  6. Wang G, He Y, Guo Q et al (2023) Continuous renal replacement therapy with the adsorptive oXiris filter may be associated with the lower 28-day mortality in sepsis: a systematic review and meta-analysis. Crit Care 27:275. https://doi.org/10.1186/s13054-023-04555-x

    Article  PubMed  PubMed Central  Google Scholar 

  7. Acharya M, Berger R, Popov A-F (2022) The role of the ADVanced Organ Support (ADVOS) system in critically ill patients with multiple organ failure. Artif Organs. https://doi.org/10.1111/aor.14188

    Article  PubMed  PubMed Central  Google Scholar 

  8. Fuhrmann V, Weber T, Roedl K et al (2020) Advanced organ support (ADVOS) in the critically ill: first clinical experience in patients with multiple organ failure. Ann Intensive Care 10:96. https://doi.org/10.1186/s13613-020-00714-3

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  9. Fuhrmann V, Ruiz P, de Garibay A, Faltlhauser A et al (2021) Registry on extracorporeal multiple organ support with the advanced organ support (ADVOS) system: 2-year interim analysis. Medicine (Baltimore). https://doi.org/10.1097/MD.0000000000024653

    Article  PubMed  Google Scholar 

  10. Allescher J, Rasch S, Wiessner JR et al (2021) Extracorporeal carbon dioxide Removal (ECCO(2) R) with the Advanced Organ Support (ADVOS) system in critically ill COVID-19 patients. Artif Organs. https://doi.org/10.1111/aor.14044

    Article  PubMed  PubMed Central  Google Scholar 

  11. Kaps L, Ahlbrand CJ, Gadban R et al (2021) Applicability and safety of discontinuous ADVanced Organ Support (ADVOS) in the treatment of patients with acute-on-chronic liver failure (ACLF) outside of intensive care. PLoS ONE 16:e0249342. https://doi.org/10.1371/journal.pone.0249342

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  12. Huber W, Henschel B, Schmid R et al (2017) First clinical experience in 14 patients treated with ADVOS: a study on feasibility, safety and efficacy of a new type of albumin dialysis. BMC Gastroenterol 17:32. https://doi.org/10.1186/s12876-017-0569-x

    Article  PubMed  PubMed Central  Google Scholar 

  13. Ruiz P, de Garibay A, Kellum JA, Honigschnabel J et al (2019) Respiratory and metabolic acidosis correction with the ADVanced Organ Support system. Intensive Care Med Exp 7:56. https://doi.org/10.1186/s40635-019-0269-7

    Article  Google Scholar 

  14. Falkensteiner C, Kortgen A, Leonhardt J et al (2020) Comparison of albumin dialysis devices molecular adsorbent recirculating system and ADVanced Organ Support system in critically ill patients with liver failure—a retrospective analysis. Ther Apher Dial. https://doi.org/10.1111/1744-9987.13533

    Article  PubMed  Google Scholar 

  15. Sommerfeld O, Neumann C, Becker J et al (2023) Extracorporeal albumin dialysis in critically ill patients with liver failure: comparison of four different devices—a retrospective analysis. Int J Artif Organs 46:481–491. https://doi.org/10.1177/03913988231191952

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  16. Erra Díaz F, Dantas E, Geffner J (2018) Unravelling the interplay between extracellular acidosis and immune cells. Mediators Inflamm 2018:1218297. https://doi.org/10.1155/2018/1218297

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  17. Shimazui T, Nakada T, Tateishi Y et al (2019) Association between serum levels of interleukin-6 on ICU admission and subsequent outcomes in critically ill patients with acute kidney injury. BMC Nephrol 20:74. https://doi.org/10.1186/s12882-019-1265-6

    Article  PubMed  PubMed Central  Google Scholar 

  18. Jarczak D, Kluge S, Nierhaus A (2024) Septic hyperinflammation—is there a role for extracorporeal blood purification techniques? Int J Mol Sci. https://doi.org/10.3390/ijms25063120

    Article  PubMed  PubMed Central  Google Scholar 

  19. Sinha P, Calfee CS (2019) Phenotypes in acute respiratory distress syndrome: moving towards precision medicine. Curr Opin Crit Care 25:12–20. https://doi.org/10.1097/MCC.0000000000000571

    Article  PubMed  PubMed Central  Google Scholar 

  20. Kaps L, Medina-Montano C, Bros M et al (2023) Comparison of inflammatory cytokine levels in hepatic and jugular veins of patients with cirrhosis. Mediators Inflamm 2023:9930902. https://doi.org/10.1155/2023/9930902

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  21. Bonavia A, Groff A, Karamchandani K et al (2018) Clinical utility of extracorporeal cytokine hemoadsorption therapy: a literature review. Blood Purif 46:337–349. https://doi.org/10.1159/000492379

    Article  CAS  PubMed  Google Scholar 

  22. Wolley M, Jardine M, Hutchison CA (2018) Exploring the clinical relevance of providing increased removal of large middle molecules. Clin J Am Soc Nephrol 13:805–814. https://doi.org/10.2215/CJN.10110917

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  23. Maduell F, Broseta JJ, Rodríguez-Espinosa D et al (2023) Efficacy and safety of the medium cut-off Elisio HX dialyzer. Blood Purif 52:68–74. https://doi.org/10.1159/000524644

    Article  CAS  PubMed  Google Scholar 

  24. Reis T, Martino F, Dias P et al (2021) Removal of middle molecules with medium cutoff dialyzer in patients on short frequent hemodialysis. Hemodial Int 25:180–187. https://doi.org/10.1111/hdi.12906

    Article  PubMed  Google Scholar 

  25. Weiner DE, Falzon L, Skoufos L et al (2020) Efficacy and safety of expanded hemodialysis with the Theranova 400 Dialyzer: a randomized controlled trial. Clin J Am Soc Nephrol 15:1310–1319. https://doi.org/10.2215/CJN.01210120

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  26. Ronco C, Clark WR (2018) Haemodialysis membranes. Nat Rev Nephrol 14:394–410. https://doi.org/10.1038/s41581-018-0002-x

    Article  CAS  PubMed  Google Scholar 

  27. Kneis C, Beck W, Boenisch O et al (2013) Elimination of middle-sized uremic solutes with high-flux and high-cut-off membranes: a randomized in vivo study. Blood Purif 36:287–294. https://doi.org/10.1159/000356224

    Article  CAS  PubMed  Google Scholar 

  28. Harm S, Schildböck C, Hartmann J (2020) Cytokine removal in extracorporeal blood purification: an in vitro study. Blood Purif 49:33–43. https://doi.org/10.1159/000502680

    Article  CAS  PubMed  Google Scholar 

  29. Pfensig C, Dominik A, Borufka L et al (2016) A new application for albumin dialysis in extracorporeal organ support: characterization of a putative interaction between human albumin and proinflammatory cytokines IL-6 and TNFα. Artif Organs 40:397–402. https://doi.org/10.1111/aor.12557

    Article  CAS  PubMed  Google Scholar 

  30. Glancey G (2021) Modeling the transfer of low molecular weight plasma proteins during hemodialysis and online hemodiafiltration. Artif Organs 45:419–426. https://doi.org/10.1111/aor.13835

    Article  CAS  PubMed  Google Scholar 

  31. Kaps L, Schleicher EM, Medina Montano C et al (2022) Influence of Advanced Organ Support (ADVOS) on cytokine levels in patients with acute-on-chronic liver failure (ACLF). J Clin Med. https://doi.org/10.3390/jcm11102782

    Article  PubMed  PubMed Central  Google Scholar 

  32. Schädler D, Pausch C, Heise D et al (2017) The effect of a novel extracorporeal cytokine hemoadsorption device on IL-6 elimination in septic patients: a randomized controlled trial. PLoS ONE 12:e0187015. https://doi.org/10.1371/journal.pone.0187015

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  33. Said EA, Al-Reesi I, Al-Shizawi N et al (2021) Defining IL-6 levels in healthy individuals: a meta-analysis. J Med Virol 93:3915–3924. https://doi.org/10.1002/jmv.26654

    Article  CAS  PubMed  Google Scholar 

  34. Heymann M, Schorer R, Putzu A (2023) The effect of CytoSorb on inflammatory markers in critically ill patients: a systematic review and meta-analysis of randomized controlled trials. Crit Care Med 51:1659–1673. https://doi.org/10.1097/CCM.0000000000006007

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  35. Hirayama Y, Oda S, Wakabayashi K et al (2011) Comparison of interleukin-6 removal properties among hemofilters consisting of varying membrane materials and surface areas: an in vitro study. Blood Purif 31:18–25. https://doi.org/10.1159/000321142

    Article  CAS  PubMed  Google Scholar 

  36. Eichhorn T, Hartmann J, Harm S et al (2017) Clearance of selected plasma cytokines with continuous veno-venous hemodialysis using ultraflux EMiC2 versus ultraflux AV1000S. Blood Purif 44:260–266. https://doi.org/10.1159/000478965

    Article  CAS  PubMed  Google Scholar 

  37. Lumlertgul N, Hall A, Camporota L et al (2021) Clearance of inflammatory cytokines in patients with septic acute kidney injury during renal replacement therapy using the EMiC2 filter (Clic-AKI study). Crit Care 25:39. https://doi.org/10.1186/s13054-021-03476-x

    Article  PubMed  PubMed Central  Google Scholar 

  38. Masakane I, Sakurai K (2018) Current approaches to middle molecule removal: room for innovation. Nephrol Dial Transplant 33:iii12–iii21. https://doi.org/10.1093/ndt/gfy224

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  39. Honore PM, Hoste E, Molnár Z et al (2019) Cytokine removal in human septic shock: where are we and where are we going? Ann Intensive Care 9:56. https://doi.org/10.1186/s13613-019-0530-y

    Article  PubMed  PubMed Central  Google Scholar 

  40. Ruiz P, de Garibay A, Honigschnabel J, Kreymann B (2018) The ADVOS Device: albumin dialysis based approach for significant and continuous removal of water soluble and protein bound toxins even at low blood flows: 36th Vicenza Course on AKI & CRRT. Blood Purif 46(163–186):P14

    Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

SH, APR and TMB participated in the study design and conceptualization. SH and APR contributed with the statistical analysis and wrote the manuscript. VB and FZ participated in data curation. TMB provided critical review with important intellectual input. All authors are accountable for all aspects of the work, and all authors read and approved the final manuscript.

Corresponding author

Correspondence to Aritz Perez Ruiz de Garibay.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

SH, APR and TB work at ADVITOS GmbH. Other authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark