Predictive validity of a novel non-invasive estimation of effective shunt fraction in critically ill patients

A Correction to this article was published on 10 July 2020

A Letter to the Editor to this article was published on 02 July 2020

This article has been updated

Abstract

Background

Accurate measurement of pulmonary oxygenation is important for classification of disease severity and quantification of outcomes in clinical studies. Currently, tension-based methods such as P/F ratio are in widespread use, but are known to be less accurate than content-based methods. However, content-based methods require invasive measurements or sophisticated equipment that are rarely used in clinical practice. We devised two new methods to infer shunt fraction from a single arterial blood gas sample: (1) a non-invasive effective shunt (ES) fraction calculated using a rearrangement of the indirect Fick equation, standard constants, and a procedural inversion of the relationship between content and tension and (2) inferred values from a database of outputs from an integrated mathematical model of gas exchange (DB). We compared the predictive validity—the accuracy of predictions of PaO2 following changes in FIO2—of each measure in a retrospective database of 78,159 arterial blood gas (ABG) results from critically ill patients.

Results

In a formal test set comprising 9,635 pairs of ABGs, the median absolute error (MAE) values for the four measures were as follows: alveolar-arterial difference, 7.30 kPa; PaO2/FIO2 ratio, 2.41 kPa; DB, 2.13 kPa; and ES, 1.88 kPa. ES performed significantly better than other measures (p < 10-10 in all comparisons). Further exploration of the DB method demonstrated that obtaining two blood gas measurements at different FIO2 provides a more precise description of pulmonary oxygenation.

Conclusions

Effective shunt can be calculated using a computationally efficient procedure using routinely collected arterial blood gas data and has better predictive validity than other analytic methods. For practical assessment of oxygenation in clinical research, ES should be used in preference to other indices. ES can be calculated at http://baillielab.net/es.

Background

Hypoxia is the defining feature of respiratory failure. Accurate quantification of pulmonary oxygenation defect is essential to determine inclusion in clinical trials, to measure outcomes in research studies, and to observe changes in lung function in a clinical setting.

In severely hypoxic patients, direct measurement of intrapulmonary shunt provides the most accurate quantification of an oxygenation defect [1]. Tension-based indices, including PaO2/FIO2 (P/F) ratio and alveolar-arterial (A-a) difference have poor agreement with intrapulmonary shunt fraction [13]. The primary limitation in tension-based indices is the marked and non-linear change in PaO2 when FIO2 is changed [4]. Brochard and colleagues demonstrated that this can be predicted from a simple mathematical model [5].

The concept of predictive validity is a mathematical reality check for a clinical measure. For a given clinical measure, predictive validity quantifies the extent to which that measure predicts an unseen event. The intent is not to predict the future, but rather to provide a rigourous, unbiased test of how well a clinical measure is describing a real entity: the assumption is that whichever measure is closest to the truth should also provide the best prediction. This approach, using mortality as the predicted event, was used in the development of consensus definitions for both acute respiratory distress syndrome (ARDS) [6] and sepsis [7].

A measure that accurately reflects the true state of a patient’s lungs should not change markedly following a change in FIO2. Therefore, the prediction of a PaO2 following a change in FIO2, assuming that the measure of the oxygenation defect remains unaltered, is a valid assessment for predictive validity.

We hypothesised that an easily understood, content-based oxygenation index may be obtainable from routinely-acquired arterial blood gas (ABG) data, without any need for additional invasive measurements. In order to assess different approaches, we quantified the predictive validity of P/F, A-a, and two new methods of estimating shunt fraction (effective shunt fraction (ES) and a database method (DB)) in a simple test: prediction of PaO2 following a change in FIO2 in a large retrospective cohort.

Methods

Data source and filtering

We used a set of 78,159 arterial blood gas samples taken between 2011 and 2016 from 6511 patients on the general intensive care unit (ICU) at the Royal Infirmary of Edinburgh. The unit admits adult patients, with predominantly emergency medical, trauma, and general surgery conditions—not elective cardiac or thoracic surgery or neurosurgery. We did not study patients who had ECMO. The samples were routine analyses: the FIO2 value was input by the clinician performing the analysis. The analysis machine was maintained by the clinical chemistry department and regularly calibrated against known standards.

To obtain sample sets in which underlying pulmonary pathology was unlikely to change substantially between samples, we limited the selection of samples to pairs of ABGs that met the following inclusion criteria: (1) taken within a 3-h window, (2) taken from a mechanically ventilated patient, (3) where the FIO2 was reduced between the first and the second sample, and (4) where alveolar ventilation was stable (change in PaCO2 < 0.3 kPa).

Derivation of effective shunt fraction

ES expresses the shunt fraction that would be required to produce a given impairment in oxygenation, that is the proportion of cardiac output that would have to shunt in order to have this effect (i.e. to produce this degree of hypoxia). In clinical practice, it will almost never be the case that a given patient has pure shunt; ES is intended to provide an intuitive and consistent quantification of oxygenation impairment.

Full details of the methods used are given in the Additional file 1. Briefly, ES was first calculated from the blood gas results as follows.

The shunt equation is usually expressed in the following way [8]:

$$ \frac{Q_{S}}{Q_{T}} = \frac{C_{c'}O_{2} - C_{a}O_{2}}{C_{c'}O_{2} - C_{v}O_{2}} $$
(1)

All of the necessary variables can be easily calculated from routine clinical measurements, with the exception of CvO2. We applied the Fick principle for oxygen uptake (see Additional file 2), in order to replace this term:

$$ \frac{Q_{S}}{Q_{T}} = \frac{C_{c'}O_{2} - C_{a}O_{2}}{C_{c'}O_{2} - C_{a}O_{2} - \frac{VO_{2}}{Q}} $$
(2)

After estimating PAO2 using the alveolar gas equation, arterial (CaO2) and end-capillary (Cc’O2) oxygen contents were derived using model equations from Dash and Bassingthwaighte [9], using measured pH and PaCO2 to estimate the PO2 at which Hb is 50% saturated (P50). Values for oxygen consumption (VO2) and cardiac output (Q) were set at single values in the physiological range (see Additional file 1).

Predictive validity

Each measure was quantified for the earlier ABG in each pair. For ES, P/F and A-a, the FIO2 and PAO2 for the second ABG was used in the rearranged equations (derived in Additional file 1) to estimate the new PaO2, using the same value for the oxygenation index under inspection.

For measurement of predictive validity in the ES method, a predicted CaO2, after change of FIO2, was calculated using FIO2 and PAO2 from the second ABG:

$$ C_{a}O_{2} = C_{c'}O_{2} - \frac{Q_{S}/Q_{T} \times VO_{2}}{Q_{T} - Q_{S}} $$
(3)

This value of CaO2 was then converted to a predicted PaO2 value using the method of Dash and Bassingthwaighte [9].

In order to minimise the noise generated (affecting all measures) by changes in ventilation or circulation, we focused our study on patients whose FIO2 was being weaned downwards. The median absolute differences between predicted and observed PaO2 across all ABG pairs were taken as the predictive validity for each measure. For the DB method, input settings for an integrated mathematical model of gas exchange were identified which matched to the first ABG. These were then extrapolated to the FIO2 and PaCO2 of the second ABG, and the mean PaO2 of all matching model runs taken to be the prediction (Fig. 1b). Further details of the mathematical model used for the DB method can be found in supplementary content (see Additional file 2).

Fig. 1
figure 1

a Boxplot showing distribution of absolute error for each measure in all samples, together with baseline distribution of pairs of ABGs in which FIO2 was unchanged (box shows mean +/ − one quartile, whiskers show range). b Range of possible FIO2-PaO2 combinations for conditions matching a single ABG result. c Range of possible results for conditions matching two ABG results at different FIO2

Research Ethics

Ethical approval was obtained from the Scotland A Research Ethics Committee [16/SS/0209].

Software and statistical analyses

All analysis was performed using Python 3.5.2 and scipy.stats version 0.18.1. A Kruskal-Wallis H test was used to determine the difference in error rate between the different measures. Mann-Whitney U tests with Bonferroni correction were used as a stringent post test for pairwise comparisons.

Results

Comparison of oxygenation measures

From the total set of 78,159 ABGs from 6511 patients, an initial test set was selected at random, containing 54,115 ABGs from 4558 patients. From this random sample, we selected a formal test set comprising 9635 pairs of ABGs, which met the criteria listed above, pairs of ABGs taken from mechanically ventilated patients within a 3-h window, where the FIO2 was reduced between the first and the second sample, and where alveolar ventilation was stable. When we compared the predicted with the measured PaO2 values in the second values of these pairs, the median absolute error (MAE) values for the four measures considered were as follows: A-a, 7.30 kPa; P/F, 2.41 kPa; DB, 2.13 kPa; and ES, 1.88 kPa. ES had significantly superior predictive validity than all other measures (Table 1).

Table 1 Pairwise comparisons between errors in oxygenation measures in test set (Mann-Whitney U test, Bonferroni correction)

Effective shunt values in this population ranged from 0 to 63% (mean 16.1%, SD 8.6%). P/F values (kPa) ranged from 6.65 to 84.7 (mean 32.2, SD 12.5). A-a values (kPa) ranged from 0 to 81.2 (mean 22.7, SD 13.6).

Validation of assumed values

Three key assumed values are required for the calculation of ES: respiratory exchange ratio (RER), cardiac output (Q), and metabolic oxygen consumption (VO2). In order to prevent bias, we optimised estimation of these assumed parameters in a training set comprising 30% of the available ABGs (n=24,044), selected at random. Varying the three values over wide ranges (RER, 0.8 to 1.1; Q, 3 to 15 l.min−1; VO2, 0.15 to 1 l.min−1) had minimal effect on the value of effective shunt.

Multiple ABGs under different conditions

Although the DB method does not perform as well as the simpler and less computationally demanding ES method, it provides an opportunity to test the effect of obtaining multiple ABGs at different FIO2. The model takes standard physiological inputs, including pure shunt fraction, V/Q heterogeneity index, cardiac output, and FIO2 and returns blood gas results at steady state. The database of model results enables us to infer the possible physiological conditions that could give rise to a given ABG result. With one ABG, a wide range of possibilities remain (Fig. 1b); a second ABG at a different FIO2 substantially constrains the range of possible physiological states that describe a given patient (Fig. 1c).

Discussion

This is the first study, to our knowledge, comparing the predictive validity of non-invasive effective shunt fraction with tension-based measures. Our observations are consistent in magnitude and direction with previous work studying changes in measures of oxygenation in human participants [13].

The poor performance of A-a difference is consistent with the report by Cane and colleagues [1], who demonstrated that A-a was the least reliable measure compared with invasive measurements of Qs/Qt. The very low predictive validity of A-a in our study, together with previous work, leads us to conclude that this measure has no role in any context.

Our study has several limitations. The ABG data itself was obtained from electronic records whereby FIO2 was entered by the treating clinician (nurse/doctor/nurse practitioner), which is a possible source of error. Importantly, this potential error applies equally to all measures of oxygenation.

To compare the integrity of each oxygenation measure alone, we assumed that baseline physiological function is not altered by the change of FIO2. However, increases in FIO2 cause absorption atelectasis [10]. We have mitigated this by restricting our analysis to pairs of ABGs in which the FIO2 was decreasing. Since the reversal of absorption atelectasis is slower than the onset [10], and there are fewer sudden changes in oxygenation in this group, we expect that restricting our analysis to weaning patients will mitigate this source of noise.

Marked changes in PaO2 may occur within a 3-h interval due to real changes in pulmonary function, for example due to recruitment, suction, diuresis, or change in posture. We therefore cannot draw any inference from the absolute value of the error in prediction, only a comparison between different methods. Noise caused by these and other factors is expected to limit the maximum possible accuracy of any prediction of PaO2. This minimum achievable error is reflected by the baseline values (Fig. 1a) showing the change in PaO2 between pairs of ABGs meeting the other selection criteria, with no change in FIO2.

There is also a significant limitation in the concept of reducing the full complexity of pulmonary oxygenation to a single numerical value. All clinical measurements are subject to this limitation—they provide summary measurements that require informed interpretation. The lung is no different from any other system in this regard. Studies using the multiple gas elimination technique (MIGET) have confirmed that lung injury leads to substantial heterogeneity in the matching of ventilation to perfusion, which causes hypoxia without pure shunt [11, 12]. ES, by design, combines these mechanisms into a single value from a three-compartment model: the amount of pure shunt that would be needed to have a given effect on oxygenation.

Since V/Q heterogeneity and shunt are separate inputs into the physiological model used to generate the database (see Additional file 1), the database approach is expected to handle this distinction better than the other measures. However, as shown in Fig. 1b, there is insufficient information in a single ABG to distinguish between shunt and V/Q heterogeneity. In contrast, with two ABGs taken at different settings of FIO2, the patient’s oxygen responsiveness is quantified, greatly restricting the range of possible values for both shunt and V/Q heterogeneity (Fig. 1c). This double FIO2 test may resolve the uncertainty in quantifying pulmonary shunt but is, at present, computationally demanding.

The striking superiority of ES in the context of critical illness may lead to an increase in clinical use. We support this, in part because the value itself is intuitive to critical care clinicians and is comparable across different health care systems and measurement units. Although it performs substantially better than other measures, it should be noted that ES is an imperfect measure and is not expected to be completely independent of extra-pulmonary factors, including FIO2, alveolar ventilation, and intracardiac shunt.

Conclusions

The effective shunt fraction, a new, non-invasive method of estimating shunt, can be calculated on any ABG result, provided the FIO2 is known. The computation is fast and simple. Hence, the method could be retrospectively applied to previous studies that hold ABG data in a machine-readable format. Whilst the simplicity of the P/F ratio will continue to make it a popular choice for clinical use, the superior predictive validity of ES makes it a better choice where accurate quantification of oxygenation defect is necessary.

Availability of data and materials

An online calculator to compute the effective shunt fraction is available at: http://baillielab.net/es Python code to calculate the effective shunt fraction is available from github: http://github.com/baillielab

Change history

  • 10 July 2020

    An amendment to this paper has been published and can be accessed via the original article.

Abbreviations

A-a:

Alveolar-arterial difference

ABG:

Arterial blood gas

ARDS:

Acute respiratory distress syndrome

CaO2 :

Arterial blood oxygen content

Cc’O2 :

End-capillary blood oxygen content

DB:

Database method

ES:

Effective shunt

FIO2; ICU:

Intensive care unit

MWu:

Mann-Whitney U test

P/F:

The ratio of PaO2 in the arterial blood to the therapeutic MAE: Median absolute error

Q:

Cardiac output

Qs/Qt:

Shunt fraction

RER:

Respiratory exchange ratio

V/Q:

Ventilation/perfusion ratio

VO2 :

Metabolic oxygen consumption

References

  1. Cane RD, Shapiro BA, Templin R, et al. (1988) Unreliability of oxygen tension-based indices in reflecting intrapulmonary shunting in critically ill patients. Crit Care Med 16:1243–5.

    Article  CAS  Google Scholar 

  2. Drummond GB, Zhong NS (1983) Inspired oxygen and oxygen transfer during artificial ventilation for respiratory failure. Br J Anaesth 55:3–13.

    Article  CAS  Google Scholar 

  3. Allardet-Servent J, Forel J-M, Roch A, et al. (2009) FIO2 and acute respiratory distress syndrome definition during lung protective ventilation. Crit Care Med 37:e4–6. https://doi.org/10.1097/CCM.0b013e31819261db.

    Article  Google Scholar 

  4. Gowda MS, Klocke RA (1997) Variability of indices of hypoxemia in adult respiratory distress syndrome. Crit Care Med 25:41–5.

    Article  CAS  Google Scholar 

  5. Aboab J, Louis B, Jonson B, et al. (2006) Relation between pao2/fio2 ratio and fio2: a mathematical description. Intensive Care Med 32:1494–7. https://doi.org/10.1007/s00134-006-0337-9.

    Article  Google Scholar 

  6. ARDS Definition Task Force, Ranieri VM, Rubenfeld GD, et al. (2012) Acute respiratory distress syndrome: the Berlin Definition. JAMA 307:2526–33. https://doi.org/10.1001/jama.2012.5669.

    Google Scholar 

  7. Shankar-Hari M, Phillips GS, Levy ML, et al. (2016) Developing a new definition and assessing new clinical criteria for septic shock: For the Third International Consensus Definitions for Sepsis and Septic Shock (Sepsis-3). JAMA 315:775–87. https://doi.org/10.1001/jama.2016.0289.

    Article  CAS  Google Scholar 

  8. Benatar SR, Hewlett AM, Nunn JF (1973) The use of iso-shunt lines for control of oxygen therapy. Br J Anaesth 45:711–8.

    Article  CAS  Google Scholar 

  9. Dash RK, Bassingthwaighte JB (2010) Erratum to: Blood hbo2 and hbco2 dissociation curves at varied o2, co2, pH, 2,3-dpg and temperature levels. Ann Biomed Eng 38:1683–701. https://doi.org/10.1007/s10439-010-9948-y.

    Article  Google Scholar 

  10. Santos C, Ferrer M, Roca J, et al. (2000) Pulmonary gas exchange response to oxygen breathing in acute lung injury. Am J Respir Crit Care Med 161:26–31. https://doi.org/10.1164/ajrccm.161.1.9902084.

    Article  CAS  Google Scholar 

  11. Duenges B, Vogt A, Bodenstein M, et al. (2009) A comparison of micropore membrane inlet mass spectrometry-derived pulmonary shunt measurement with riley shunt in a porcine model. Anesth Analg 109:1831–5. https://doi.org/10.1213/ANE.0b013e3181bbc401.

    Article  Google Scholar 

  12. Wagner PD (2008) The multiple inert gas elimination technique (MIGET). Intensive Care Med 34:994–1001. https://doi.org/10.1007/s00134-008-1108-6.

    Article  Google Scholar 

Download references

Acknowledgments

This research was supported by The University of Edinburgh and NHS Lothian.

Funding

JKB is grateful to acknowledge funding support from a Wellcome Trust Intermediate Clinical Fellowship (103258/Z/13/Z) and a Wellcome-Beit Prize (103258/Z/13/A), BBSRC Institute Strategic Programme Grant to the Roslin Institute, and the UK Intensive Care Society. A.B. is grateful to acknowledge funding from Edinburgh Clinical Academic Track and funding from the Wellcome Trust (204979/Z/16/Z).

Author information

Authors and Affiliations

Authors

Contributions

JKB and EMC designed the study and conducted the computational analysis. JKB and AB wrote the computational model of gas exchange. GBD contributed to the conception of hypothesis and interpretation of the results. JKB, EMC, and AB wrote the manuscript with assistance from GBD. All authors commented on or contributed to the final manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to J Kenneth Baillie.

Ethics declarations

Ethics approval and consent to participate

Ethical approval was obtained from the Scotland A Research Ethics Committee [16/SS/0209].

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Additional files

Additional file 1

Supplementary information. Includes the derivation of oxygenation measures and method of predicting PaO2, as well as optimisation of assumed variables using the test dataset. (PDF 227 kb)

Additional file 2

Integrated model of gas exchange. Structure of an integrated computational model of oxygen delivery which was used to generate a database of model results for inferring the possible physiological conditions that could give to rise to any given ABG result (“DB” method in the main manuscript). (PDF 80 kb)

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Chang, E.M., Bretherick, A., Drummond, G.B. et al. Predictive validity of a novel non-invasive estimation of effective shunt fraction in critically ill patients. ICMx 7, 49 (2019). https://doi.org/10.1186/s40635-019-0262-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40635-019-0262-1

Keywords