Low flow rate alters haemostatic parameters in an ex-vivo extracorporeal membrane oxygenation circuit

Abstract

Background

Extracorporeal membrane oxygenation (ECMO) is a life-saving modality used to manage cardiopulmonary failure refractory to conventional medical and surgical therapies. Despite advances in ECMO equipment, bleeding and thrombosis remain significant complications. While the flow rate for ECMO support is well recognized, less is known about the minimum-rate requirements and haemostasis. We investigated the relationship between different ECMO flow rates, and their effect on haemolysis and coagulation.

Methods

Ten ex-vivo ECMO circuits were tested using donated, < 24-h-old human whole blood, with two flow rates: high-flow at 4 L/min (normal adult cardiac output; n = 5) and low-flow at 1.5 L/min (weaning; n = 5). Serial blood samples were taken for analysis of haemolysis, von Willebrand factor (vWF) multimers by immunoblotting, rotational thromboelastometry, platelet aggregometry, flow cytometry and routine coagulation laboratory tests.

Results

Low-flow rates increased haemolysis after 2 h (p = 0.02), 4 h (p = 0.02) and 6 h (p = 0.02) and the loss of high-molecular-weight vWF multimers (p = 0.01), while reducing ristocetin-induced platelet aggregation (p = 0.0002). Additionally, clot formation times were prolonged (p = 0.006), with a corresponding decrease in maximum clot firmness (p = 0.006).

Conclusions

In an ex-vivo model of ECMO, low-flow rate (1.5 L/min) altered haemostatic parameters compared to high-flow (4 L/min). Observed differences in haemolysis, ristocetin-induced platelet aggregation, high-molecular-weight vWF multimers and clot formation time suggest an increased risk of bleeding complications. Since patients are often on ECMO for protracted periods, extended-duration studies are required to characterise long-term ECMO-induced haemostatic changes.

Background

Extracorporeal membrane oxygenation (ECMO) is a potentially life-saving modality that is used to provide temporary cardiopulmonary (venoarterial or VA) or pulmonary (venovenous or VV) support for critically ill patients. Despite this modality’s benefits, haemostatic alterations during ECMO are common and involve both bleeding and thrombotic events [1]. Considerable progress has been made in both circuit design and cannulation techniques; however, interaction between blood and the circuit’s artificial surface continues to trigger inflammation and disrupt normal haemostasis [2]. The result is hypercoagulability, counteracted by excessive fibrinolysis and the subsequent consumption of clotting factors, thrombocytopenia and impaired platelet function [3]. To maintain adequate oxygen delivery and CO2 removal, a typical ECMO flow rate for an adult is between 60 to 80 mL/kg/min (4 to 6 L/min), which is gradually reduced to less than 30% of that total for weaning [4].

Recommended flow rates for ECMO support have been well-documented [5]; however, less is known about how minimum blood flow rates affect haemolysis and coagulation. This is especially relevant during ECMO weaning. While reducing ECMO blood flow rates is necessary for safe weaning from VA ECMO, weaning practices for VV ECMO show wide variability [6]. The strategies reported by experienced ECMO centres have included (i) maintaining blood flow rates while weaning fresh (or sweep) gas flow; (ii) reducing ECMO blood pump flows while either maintaining fresh gas flows; or (iii) a combination of the two strategies. These differences result from divergent expert opinions and sparsity of evidence regarding the merits of reducing pump flows to lessen shear stress to the blood elements, versus an increased risk of thrombosis. Furthermore, the effect of low-flow also has implications for extracorporeal carbon dioxide removal (ECCO2R), where low- and ultra-low-flow (to as low as 250 mL/min) is considered desirable for reducing the invasiveness of extracorporeal supportive therapy in acute respiratory distress syndrome and refractory hypercapnia [7]. Blood flow through ECCO2R circuits is achieved either by using the arteriovenous gradient of the patient (pumpless arterio-venous) or using centrifugal or diagonal low-impact flow pumps (veno-venous) [8]. Thus, a more complete understanding of this process will help to optimise both weaning and anti-coagulation management strategies and provide justification for reducing or aiming for low-flow extracorporeal support.

Vascular access cannulae and centrifugal pumps, used for ECMO support, can alter fluid dynamics, often producing high shear stress and turbulent flow [9, 10]. This change in haemodynamics can cause damage to circulating blood cells and haemostatic proteins, and reduce the binding of von Willebrand factor (vWF) to collagen and platelets, resulting in a state of thrombosis, fibrinolysis and compromised platelet function [11]. Enhanced vWF cleavage and the successive loss of its high molecular weight (HMW) multimers is believed to be one of the main causes of bleeding complications in ventricular assist device patients [12]. HMW vWF multimer bands are also lost in patients undergoing ECMO support and acquired von Willebrand syndrome has been linked to the loss of platelet adhesion receptors in these patients [10, 13].

The aim of the current study was to investigate the impact of ECMO when blood circulates at high- (standard practice to match cardiac output) versus low-flow (used while patients are weaned off ECMO). We hypothesised that, relative to low-flow rates, a higher flow rate would increase haemolysis, prolong clot formation times, reduce platelet aggregation and lead to the increased consumption of coagulation factors.

Materials and methods

Experimental design and ECMO setup

All methods were approved and performed in accordance with the guidelines and regulations set forth by the Metro North Ethics Committee (HREC/16/QPCH/320). Informed consent for study participation was obtained from the Australian Red Cross Blood Service (17-11QLD-05). Blood circulation loops were developed based upon ASTM F 1841-97 Standard Practice for the Assessment of Hemolysis in Continuous Flow Blood Pumps [14]. Ten ex vivo ECMO circuits (Permanent Life Support System (PLS), Maquet CP, Rastatt, Germany) were tested (Fig. 1). The PLS Set consists of an oxygenator and ROTAFLOW centrifugal pump, both incorporated into a tubing set, with a tip-to-tip BIOLINE (recombinant human albumin and heparin) coating. Donated human whole blood units (Australian Red Cross Blood Service, Kelvin Grove, Queensland) were collected into top and bottom bags with a built in filter, containing CPD (citrate phosphate dextrose; 66.5 mL; Macopharma, Mouvaux, France). Bags were stored at 4 °C and used within 24 h, with blood subsequently circulated at two flow rates: 4 L/min (normal cardiac output; n = 5) and 1.5 L/min (weaning; n = 5) (total blood volume in the circuit, 420 ± 50 mL). The PLS system was initially primed with 0.9% sodium chloride before whole blood was introduced directly into the circuit from the blood bag. ROTAFLOW centrifugal pump speed was adjusted to 2000 (average 2033 ± 13 rpm) for circulation, and CO2 enhanced gas (5% CO2, 21% O2, 74% N2) was added at 2.5 L/min and heat exchange started. Heparin (400 IU; Pfizer, NSW, Australia), calcium chloride (3.33 mg/mL; Phebra, NSW, Australia) and sodium bicarbonate (0.016 mmol/mL; Phebra) were subsequently added. Once flow had stabilized, a vice-grip clamp was used in both high- and low-flow conditions. The vice-grip was tightened to adjust the blood flow rate to 4 L/min for high-flow experiments. For low-flow experiments, the centrifugal pump speed was reduced (average 898 ± 13 rpm) until a flow rate of 1.5 L/min was achieved. Inlet and outlet pressures between the membrane oxygenator were monitored with silicone strain-gauge pressure transducers (PX181B-015C5V; Omega Engineering, CT, USA) and adjusted to 100–150 mmHg using SAGM solution (saline-adenine-glucose-mannitol, MacoPharma, Australia). Calcium (> 2.5 mmol/L), glucose (19.2 ± 4.1 mmol/L) and pH (7.36 ± 0.05) levels were monitored throughout the experiment with rotational thromboelastometry (ROTEM) used to monitor anti-coagulation requirements.

Fig. 1
figure 1

Schematic diagram of ex-vivo ECMO model setup

Blood gas and sample collection

Blood was analysed for electrolytes, glucose and pH using a handheld point-of-care system (Abbott i-STAT, Chicago, ILL, USA). Samples were collected at baseline, 1, 2, 4 and 6 h for full blood examination (FBE), flow cytometry and ROTEM® analysis, as well as for Multiplate® whole blood platelet function testing. After blood sampling, an equivalent volume of SAGM was injected back into the system to maintain pressure at 100–150 mmHg. A full blood examination was determined using the AcT diff™ haematology analyser (Beckman Coulter Australia Pty Ltd, NSW, Australia) to assess the white cell count (WCC), haemoglobin (Hb) level and platelet count (PLT). Samples for vWF multimers and routine/specialized coagulation tests (including ELISA’s and Luminex assays) were centrifuged twice (15 min, 4 °C, 3000×g) to obtain platelet-poor plasma (PPP) and subsequently stored at − 80 °C, until further analysis. Samples for haemolysis were centrifuged once (15 min, 4 °C, 3000×g).

Haemolysis assay

Plasma samples (100 μL) were diluted with 1 mL 0.1% Na2CO3 solution and absorbance measured using a UV/visible spectrophotometer (SmartSpec 3000, Bio-Rad, California, USA) at three wavelengths: 380, 415 and 450 nm. Plasma-free haemoglobin (pfHb) level was calculated using the equation below [15].

$$ \mathrm{pfHb}\ \left(\frac{\mathrm{mg}}{\mathrm{dL}}\right)=\left(167.2\ \mathrm{x}\ A415-83.6\ \mathrm{x}\ A380-83.6\ \mathrm{x}\ A450\right)\ \mathrm{x}\frac{1}{10}\ \mathrm{x}\ \left(1/\frac{\mathrm{Vol}\ \mathrm{plasma}}{\mathrm{Vol}\ {\mathrm{Na}}_2{\mathrm{CO}}_3}\right) $$

Whole blood platelet function (Multiplate®)

Whole blood platelet aggregometry was measured by change in impedance, using the Multiplate® 5.0—platelet function analyser (Haemoview Diagnostics, Brisbane, Australia). Hirudin-anticoagulated blood was tested using ristocetin (RISTO; 0.77 mg/mL), adenosine diphosphate (ADP; 6.5 μM) and thrombin receptor activating peptide-6 (TRAP-6; 32 μM) agonists. Data was expressed as the area under the aggregation curve (AUC).

Immunoblotting for high and low molecular weight vWF

Electrophoresis was performed as per previously published methodology [16]. Briefly, samples were run on a gel using high gelling temperature SeaKem agarose (Lonza, Basel, Switzerland) and transferred to a PVDF membrane (Merck KGaA, Darmstadt, Germany). Membranes were blocked using 5% skim milk in TBS/Tween and probed with horseradish peroxidase conjugated polyclonal rabbit anti-human vWF (1:1000) (DAKO, Glostrup, Denmark). Membranes were washed and developed with enhanced chemiluminescence (ECL; Amersham Bioscience, NJ, USA), and band density visualized with the ImageQuant LAS 4000 (GE Healthcare, NJ, USA). Protein expression was quantified using ImageJ (National Institute of Health) and data normalised to baseline measurements.

Flow cytometry staining

Whole blood collected at baseline, 1 and 6 h was stained with the following monoclonal antibodies (Becton Dickinson (BD) Biosciences, California, USA): fluorescein isothiocyanate (FITC) conjugated CD41a, Brilliant violet (BV)-421 conjugated P-selectin and allophycocyanin (APC) conjugated CD42b (15 min; room temperature). Red blood cells were then lysed (FACS Lysing Solution; BD Biosciences) and samples analysed using a Fortessa X-20 flow cytometer (BD Biosciences). Platelets were first identified based upon forward and side scatter, and further confirmed by gating CD41a positive cells. The P-selectin and CD42b mean fluorescent intensity (MFI) of each sample was analysed using FCS Expression 6 (De Novo Software, California, USA).

Specialized coagulation tests

The levels of factors XII and XIII, and antithrombin (AT) III were measured in plasma using a standard sandwich ELISA (Abcam, VIC, Australia). Plasma specimens also were analysed using custom Luminex Magpix-based assays (R&D Systems, MN, USA), in accordance with manufacturer instructions. Analytes measured included ADAMTS13, P-selectin, d-dimer and Protein C. Concentrations were calculated using a five-parameter logistic standard curve, corrected for background readings.

Rotational thromboelastometry (ROTEM®) and routine coagulation tests

Whole blood clot formation profiles were recorded by ROTEM® Thromboelastometry (Haemoview Diagnostics, Brisbane, Australia) with HEPTEM (contact factor-initiated coagulation with heparinase), EXTEM (thromboplastin-initiated coagulation) and FIBTEM (thromboplastin-initiated coagulation with the platelet inhibitor cytochalasin D) activating reagents, in accordance with manufacturer instructions. Parameters evaluated included clotting time (CT), clot formation time (CFT) and maximum clot firmness (MCF). Prothrombin time (PT) was measured on the Stago STAR Evolution analyser (Diagnostica Stago, Doncaster, VIC, Australia), following manufacturer instructions.

Statistical analysis

Each data set’s distribution was assessed for normality by inspecting the corresponding histogram and pp-plot, and using a Kolmogorov-Smirnov test. Two-way repeated-measures analysis of variance (RM-ANOVA) was used to compare differences between high-flow and low-flow over time for all investigated parameters, except haemolysis and vWF multimers. When simple main effects and/or any interaction between time and condition was significant, a Tukey’s post-hoc test was performed for multiple comparisons. For haemolysis and vWF multimers, the assumption of homogeneity of variances was not met; consequently, Kruskal-Wallis tests were used to compare high-flow or low-flow over time. Measurements between the two experimental groups were compared with Mann-Whitney tests for time point. All analyses considered potential confounders; this included SAGM dilution. All statistical analysis was performed using Statistica, version 13.2 (TIBCO Software Inc, Palo Alto, CA, USA). A p value ≤ 0.05 was considered statistically significant.

Results

Full blood examination

Haematological parameters were examined at selected time points (Table 1). Baseline measurements did not differ significantly across the experimental groups. WCC, Hb, and PLTs remained constant throughout the experiment, with no significant difference over time or between high- and low-flow time points.

Table 1 Haemostatic parameters by experimental group at selected time points

Haemolysis

Haemolysis levels were significantly increased over time from baseline to 6 hours, both with high- (p = 0.01) and low-flow (p = 0.002, Fig. 2). Significant differences between the two flow rates also were evident after 2 h (p = 0.02), 4 h (p = 0.02) and 6 h (p = 0.02) with greater haemolysis during low flow.

Fig. 2
figure 2

Time-dependent haemolysis with different blood flow regimens. Haemolysis levels were significantly increased over time from baseline to 6 h with both high and low-flow rates. Significant differences also were evident between high- and low-flow at 2, 4 and 6 h. Data are presented as mean ± SEM. **p < 0.01 from baseline to 6 h; #p < 0.05 high vs. low flow. n = 5 high flow, n = 4 low flow

Platelet function (Multiplate®)

ADP- and TRAP-induced platelet aggregation were both significantly decreased over time from baseline to 6 h, with high- (p = 0.02 and p = 0.0008, respectively) and low-flow (both p = 0.0002, Fig. 3a, b). RISTO-induced platelet aggregation was significantly decreased over time from baseline to 6 h with low-flow only (p = 0.0002, Fig. 3c). There was no significant difference between high- and low-flow at individual time points.

Fig. 3
figure 3

Platelet aggregometry function measured at discrete time points. ADP- (a) and TRAP-induced (b) platelet aggregation area under the curve (AUC) was significantly decreased over time from baseline to 6 h with both high and low flow. RISTO-induced platelet aggregation (c) was significantly decreased over time from baseline to 6 h with low flow only. Data are presented as mean ± SEM. *p < 0.05; ***p < 0.001 from baseline to 6 h. n = 5 high flow, n = 5 low flow

High molecular weight vWF degradation

The levels of HMW vWF multimers were significantly reduced over time from baseline to 6 h with low-flow only (p = 0.01, Fig. 4). Significant differences between high- and low-flow also were evident after 6 h (p = 0.03) with a loss of HMW vWF multimers during low-flow. There was no difference in low molecular weight vWF multimers over time or between the two flow dynamics.

Fig. 4
figure 4

Measurement of von Willebrand factor (vWF) multimers by immunoblotting in whole blood subjected to high and low flow conditions. Representative immunoblot image from two gels run at high flow and low flow (a) with loss of high-molecular-weight (HMW) vWF bands (analysed by densitometry). Data normalized to total vWF for HMW multimers (b) and LMW multimers (c). Data are presented as mean ± SEM. **p < 0.01 from baseline to 6 h; #p < 0.05 high vs low flow. n = 5 high flow, n = 4 low flow

Flow cytometry and specialized coagulation tests

No changes were detected in the MFI of P-selectin or CD42b on platelets with either of the different flow rates when assessed by flow cytometry. However, plasma levels of P-selectin were increased from baseline to 6 h with both high- (p = 0.0002) and low-flow (p = 0.02, Fig. 5a). Protein C plasma levels were significantly increased from baseline at 6 h with high flow (p = 0.039, Fig. 5b), while levels of d-dimer were increased at 6 h with low-flow (p = 0.003, Fig. 5c). There were no changes in the plasma levels of FXII, FXIII, ATIII, or ADAMTS13 at the different flow rates.

Fig. 5
figure 5

Measurement of specialized coagulation markers. Plasma levels of P-selectin (a) were increased over time with both high and low flow. The levels of Protein C (b) were increased with high flow, while d-dimer (c) levels were increased with low flow. Data are presented as mean ± SEM. *p < 0.05; **p < 0.01; ***p < 0.001 from baseline to 6 h. n = 5 high flow, n = 5 low flow

Thromboelastometry and routine coagulation tests

Low flow resulted in prolongation of HEPTEM-CFT (p = 0.006, Fig. 6a); and overall clot quality, as measured by the HEPTEM-MCF, was lower over time from baseline to 6 h (p = 0.006, Fig. 6b). There was no difference between the individual time points during low and high flow. Additionally, EXTEM-MCF was reduced over time from baseline to 6 h with both high (p = 0.0002) and low flow (p = 0.0002, Fig. 6c). PT and FIBTEM remained unchanged across both flow conditions.

Fig. 6
figure 6

The effect of high vs. low flow on HEPTEM and EXTEM parameters. Low flow resulted in a prolonged HEPTEM-CFT (a) with a corresponding decrease in MCF (b). EXTEM-MCF was lower with both high and low flow conditions (c). Data are presented as mean ± SEM. **p < 0.01; ***p < 0.001 from baseline to 6 h. n = 5 high flow, n = 5 low flow

Discussion

We sought to determine how different flow rates in an ECMO circuit would affect haemolysis and coagulation parameters. Recommended flow rates for ECMO support are well-recognized [5]. On the other hand, lower flow rates and how they alter coagulation have not been thoroughly explored. Typical ECMO flow rates for an adult are between 60 and 80 mL/kg/min (4 to 6 L/min) and are gradually reduced to less than 30% of that total for weaning [4]. However, weaning strategies can differ between ECMO centres [6]. We chose 4 L/min (high-flow) and 1.5 L/min (low-flow) to reflect common practices. Interestingly, we found that a low-flow rate increased haemolysis and the loss of high molecular weight vWF multimers, while it also reduced RISTO-induced platelet aggregation. Clot formation times were also prolonged, with a corresponding decrease in maximum clot firmness.

Haemolysis is a common complication with centrifugal pumps, potentially resulting from a number of factors, including increased shear stress and pressure change across the oxygenator [17]. Increased pump speed is also related to the development of haemolysis [18], as such, our discovery that plasma-free haemoglobin levels are higher with low- compared to high-flow were surprising. Previously published data comparing low-flow rates in ECMO are sparse, with studies in neonates revealing no significant elevation in haemolysis [19, 20]. Cannula size has also been shown to affect shear stress and influence packed red blood cell transfusion [21]. We speculate that haemolysis is influenced by ongoing clotting in the circuit. Computer simulations with ventricular assist devices have identified increased platelet deposition and thrombus growth with flow rates of 2 L/min versus 4.5 L/min [22].

Platelets play a key role initiating normal haemostasis and pathological thrombosis. Several studies have shown that ECMO may lead to thrombocytopenia through platelet consumption [23, 24], though Abrams and colleagues found no relationship between ECMO duration and the occurrence of severe thrombocytopenia, suggesting that platelet counts might not fully reflect bleeding risk [25]. Consistent with the latter study, we found no significant change in platelet numbers. While platelet numbers were unaffected by changes in ECMO flow, studies continue to document the impact of high shear stress on platelet functionality, including impaired responses to ADP in adults [26] and children [27] on ECMO. In our study, ADP- and TRAP-induced platelet aggregation decreased over time with both high- and low-flow regimes, while RISTO-induced aggregation was reduced during low-flow only. Although our study failed to reveal any corresponding change in the number of Gp1bα or P-selectin receptors on platelets, plasma levels of P-selectin increased over time with both flow regimens. P-selectin belongs to the selectin family of cell adhesion molecules and supports the initial tethering of leukocytes to activated endothelial cells and platelets [28]. The combined loss of platelet reactivity, in the face of increased plasma P-selectin, suggests ECMO-induced platelet activation and fatigue. ECMO has been shown to increase P-selectin levels at 4 h in neonates, activating platelets whilst impairing their ability to aggregate [29].

Increased shear stress, as seen with ventricular assist devices and aortic stenosis, has been reported to cause alterations in the configuration of vWF and reduce the levels of HMW multimers [30]. The multimers of vWF uncoil under high shear stress, exposing their binding sites to the platelet receptor Gp1bα and, thus, stimulating activation [31]. Interestingly, our study revealed degradation of HMW vWF multimers with low-flow, while high-flow maintained similar levels of HMW and low-molecular-weight (LMW) multimers. These changes indicate that HMW vWF multimers are being cleaved into smaller fragments [32, 33] with low-flow. In our study, plasma levels of ADAMTS13, the vWF-cleaving protease, did not change over time. There is potentially less damage to HMW multimers with high flow, since contact time with the membrane is reduced. Blood is exposed to a large surface area with the Quadrox PLS-i (1.8 m2), and 1.5 L/min might not provide enough coverage, leading to areas of stagnation. We speculate that low pump flow leads to localised stasis and may precipitate more clot formation in the circuit oxygenator, as observed in computational modelling studies [34, 35].

Our thromboelastometry measurements of the intrinsic pathway demonstrated that low-flow prolonged the HEPTEM-CFT, associated with a corresponding decrease in MCF, consistent with thrombocytopenia/platelet dysfunction. This suggests that low-flow rate affects both the initiation and propagation phase of coagulation. With respect to the extrinsic pathway, PT was unchanged while EXTEM-MCF was reduced with both high- and low-flow, implying that changes to the extrinsic pathway may be independent of flow dynamics. This coagulation derangement may be a dilutional effect from fluid replacement.

Levels of the naturally occurring anticoagulant protein C were elevated with high-flow. This is consistent with studies in CPB that have demonstrated increased levels of activated protein C during and after bypass [36]. The detection of d-dimers, or fibrin degradation products generated by reactive fibrinolysis, with low flow, is indicative of coagulation activity. However, elevated d-dimer levels can also be seen with clinical conditions, such as inflammation, or generated by the foreign surfaces inherent with extracorporeal organ replacement therapies [37]. An increase in d-dimer concentration may, therefore, be a predictor of clot formation in the membrane oxygenator at low flow [37, 38].

Our study has a number of limitations. First, this was a study of short duration, which might not elucidate the longer-term impact of ECMO that occurs in the real-life clinical scenario. Second, we only assessed one type of oxygenator and two different flow rates (1.5 and 4 L/min), as the Quadrox PLS-i has a recommended blood flow rating of 0.5–7 L/min. Thirdly, the biological component, including any endothelial interaction, was excluded. Finally, the effect of cannula size on haemostatic changes was not considered, including the effect of shear stress induced by the edges of the venous cannulas. Future studies would focus on animal models of ECMO and assess multiple types of oxygenators at different flow rates.

Conclusions

Current management of lower ECMO blood flows balances the perceived benefits of reduced shear stress to the blood elements, and therefore reduced haemolysis, with an increased risk of thrombosis. However, our ex-vivo model enabled us to demonstrate that low-flow (1.5 L/min) during the initial hours of ECMO leads to increased haemolysis, decreased platelet aggregation, and the loss of HMW multimers. These changes to haemostasis may be due to areas of stagnation or turbulence within the oxygenator, extending the contact time between the blood and the circuit’s artificial surface. Our results suggest that there might be an ideal minimum flow rate for each oxygenator, and that haemostatic changes could result from deviations outside the optimal range. Longer-duration studies are required to characterise ECMO-induced haemostatic changes beyond 6 h with different flow rates.

Availability of data and materials

The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

ADP:

Adenosine diphosphate

APC:

Allophycocyanin

AT:

Antithrombin

AUC:

Area under the aggregation curve

BV:

Brilliant violet

CFT:

Clot formation time

CT:

Clotting time

ECCO2R:

Extracorporeal carbon dioxide removal

ECMO:

Extracorporeal membrane oxygenation

FBE:

Full blood examination

FITC:

Fluorescein isothiocyanate

Hb:

Haemoglobin

HMW:

High molecular weight

LMW:

Low molecular weight

MCF:

Maximum clot firmness

PLT:

Platelet

PT:

Prothrombin time

RISTO:

Risocetin

ROTEM:

Rotational thromboelastometry

SAGM:

Saline-adenine-glucose-mannitol

TRAP-6:

Thrombin receptor activating peptide-6

VA:

Venoarterial

VV:

Venovenous

vWF:

von Willebrand factor

WCC:

White cell count

References

  1. Thiagarajan RR, Barbaro RP, Rycus PT et al (2017) Extracorporeal life support organization registry international report 2016. ASAIO J 63(1):60–67

    Article  Google Scholar 

  2. Millar JE, Fanning JP, McDonald CI et al (2016) The inflammatory response to extracorporeal membrane oxygenation (ECMO): a review of the pathophysiology. Crit Care 20(1):387

    Article  Google Scholar 

  3. Oliver WC (2009) Anticoagulation and coagulation management for ECMO. Semin Cardiothorac Vasc Anesth 13(3):154–175

    Article  Google Scholar 

  4. ELSO (2017) ELSO Guidelines General v1.4. In: ELSO General guidelines for all ECLS Cases http://www.elso.org/resources/guidelines.aspx

    Google Scholar 

  5. Di Nardo M, Vercaemst L, Swol J et al (2018) A narrative review of the technical standards for extracorporeal life support devices (pumps and oxygenators) in Europe. Perfusion 33(7):553–561

    Article  Google Scholar 

  6. Malfertheiner MV, Broman LM, Belliato M et al (2017) Management strategies in venovenous extracorporeal membrane oxygenation: a retrospective comparison from five European centres. Crit Care Resusc 19(Suppl 1):76–81

    CAS  PubMed  Google Scholar 

  7. Combes A, Pesenti A, Ranieri VM (2017) Fifty years of research in ARDS. Is extracorporeal circulation the future of acute respiratory distress syndrome management? Am J Resp Crit Care Med 195(9):1161–1170

    Article  Google Scholar 

  8. Pettenuzzo T, Fan E, Del Sorbo L (2018) Extracorporeal carbon dioxide removal in acute exacerbations of chronic obstructive pulmonary disease. Ann Transl Med 6(2):31

    Article  Google Scholar 

  9. De Bartolo C, Nigro A, Fragomeni G et al (2011) Numerical and experimental flow analysis of the Wang-Zwische double-lumen cannula. ASAIO J 57(4):318–327

    Article  Google Scholar 

  10. Tauber H, Ott H, Streif W et al (2015) Extracorporeal membrane oxygenation induces short-term loss of high-molecular-weight von Willebrand factor multimers. Anesth Analg 120(4):730–736

    Article  CAS  Google Scholar 

  11. Heilmann C, Geisen U, Beyersdorf F et al (2012) Acquired von Willebrand syndrome in patients with extracorporeal life support (ECLS). Intensive Care Med 38(1):62–68

    Article  Google Scholar 

  12. Nascimbene A, Neelamegham S, Frazier OH et al (2016) Acquired von Willebrand syndrome associated with left ventricular assist device. Blood 127(25):3133–3141

    Article  CAS  Google Scholar 

  13. Lukito P, Wong A, Jing J et al (2016) Mechanical circulatory support is associated with loss of platelet receptors glycoprotein Ibalpha and glycoprotein VI. J Thromb Haemost 14(11):2253–2260

    Article  CAS  Google Scholar 

  14. F1841-97 (2013) Standard practice for assessment of hemolysis in continuous flow blood pumps. ASTM International, West Conshohocken

    Google Scholar 

  15. Chan CH, Hilton A, Foster G et al (2012) Reevaluation of the Harboe assay as a standardized method of assessment for the hemolytic performance of ventricular assist devices. Artif Organs 36(8):724–730

    Article  Google Scholar 

  16. Chan CHH, Pieper IL, Robinson CR et al (2017) Shear Stress-induced total blood trauma in multiple species. Artif Organs 41(10):934–947

    Article  CAS  Google Scholar 

  17. Kawahito S, Maeda T, Motomura T et al (2002) Hemolytic characteristics of oxygenators during clinical extracorporeal membrane oxygenation. ASAIO J 48(6):636–639

    Article  CAS  Google Scholar 

  18. Toomasian JM, Bartlett RH (2011) Hemolysis and ECMO pumps in the 21st Century. Perfusion 26(1):5–6

    Article  Google Scholar 

  19. Meyer AD, Wiles AA, Rivera O et al (2012) Hemolytic and thrombocytopathic characteristics of extracorporeal membrane oxygenation systems at simulated flow rate for neonates. Pediatr Crit Care Med 13(4):e255–e261

    Article  Google Scholar 

  20. Moon YS, Ohtsubo S, Gomez MR et al (1996) Comparison of Centrifugal and roller pump hemolysis rates at low flow. Artif Organs 20(5):579–581

    Article  CAS  Google Scholar 

  21. Martucci G, Panarello G, Occhipinti G et al (2018) Impact of cannula design on packed red blood cell transfusions: technical advancement to improve outcomes in extracorporeal membrane oxygenation. J Thorac Dis 10(10):5813–5821

    Article  Google Scholar 

  22. Wu WT, Yang F, Wu J et al (2016) High fidelity computational simulation of thrombus formation in Thoratec HeartMate II continuous flow ventricular assist device. Sci Rep 6:38025

    Article  CAS  Google Scholar 

  23. Malfertheiner MV, Philipp A, Lubnow M et al (2015) Hemostatic changes during extracorporeal membrane oxygenation: a prospective randomized clinical trial comparing three different extracorporeal membrane oxygenation systems. Crit Care Med 44(4):747–754

    Google Scholar 

  24. Panigada M, Artoni A, Passamonti SM et al (2016) Hemostasis changes during veno-venous extracorporeal membrane oxygenation for respiratory support in adults. Minerva Anestesiol 82(2):170–179

    PubMed  Google Scholar 

  25. Abrams D, Baldwin MR, Champion M et al (2016) Thrombocytopenia and extracorporeal membrane oxygenation in adults with acute respiratory failure: a cohort study. Intensive Care Med 42(5):844–852

    Article  CAS  Google Scholar 

  26. Mutlak H, Reyher C, Meybohm P et al (2015) Multiple electrode aggregometry for the assessment of acquired platelet dysfunctions during extracorporeal circulation. Thorac Cardiovasc Surg 63(1):21–27

    Article  Google Scholar 

  27. Saini A, Hartman ME, Gage BF et al (2015) Incidence of platelet dysfunction by thromboelastography-platelet mapping in children supported with ECMO: a pilot retrospective study. Front Pediatr 3:116

    PubMed  Google Scholar 

  28. McEver RP (1995) Regulation of function and expression of P-selectin. Agents Actions Suppl:47117–47119

  29. Cheung PY, Sawicki G, Salas E et al (2000) The mechanisms of platelet dysfunction during extracorporeal membrane oxygenation in critically ill neonates. Crit Care Med 28(7):2584–2590

    Article  CAS  Google Scholar 

  30. Ruggeri ZM, Orje JN, Habermann R et al (2006) Activation-independent platelet adhesion and aggregation under elevated shear stress. Blood 108(6):1903–1910

    Article  CAS  Google Scholar 

  31. Schneider SW, Nuschele S, Wixforth A et al (2007) Shear-induced unfolding triggers adhesion of von Willebrand factor fibers. Proc Natl Acad Sci U S A 104(19):7899–7903

    Article  CAS  Google Scholar 

  32. Bartoli CR, Restle DJ, Zhang DM et al (2015) Pathologic von Willebrand factor degradation with a left ventricular assist device occurs via two distinct mechanisms: mechanical demolition and enzymatic cleavage. J Thorac Cardiovasc Surg 149(1):281–289

    Article  CAS  Google Scholar 

  33. Tsai HM, Sussman II, Nagel RL (1994) Shear stress enhances the proteolysis of von Willebrand factor in normal plasma. Blood 83(8):2171–2179

    CAS  PubMed  Google Scholar 

  34. Funakubo A, Taga I, McGillicuddy JW et al (2003) Flow vectorial analysis in an artificial implantable lung. ASAIO J 49(4):383–387

    PubMed  Google Scholar 

  35. Gartner MJ, Wilhelm CR, Gage KL et al (2000) Modeling flow effects on thrombotic deposition in a membrane oxygenator. Artif Organs 24(1):29–36

    Article  CAS  Google Scholar 

  36. Petaja J, Pesonen E, Fernandez JA et al (1999) Cardiopulmonary bypass and activation of antithrombotic plasma protein C. J Thorac Cardiovasc Surg 118(3):422–429 discussion 429-431

    Article  CAS  Google Scholar 

  37. Lubnow M, Philipp A, Dornia C et al (2014) D-dimers as an early marker for oxygenator exchange in extracorporeal membrane oxygenation. J Crit Care 29(3):473 e471–473 e475

    Article  Google Scholar 

  38. Dornia C, Philipp A, Bauer S et al (2015) D-dimers are a predictor of clot volume inside membrane oxygenators during extracorporeal membrane oxygenation. Artif Organs 39(9):782–787

    Article  CAS  Google Scholar 

Download references

Acknowledgements

The authors thank Valentine Oqda for excellent technical assistance; the Australian Red Cross Blood Service for the provision of blood products, acknowledging that the Australian government funds the Australian Red Cross Blood Service to provide blood, blood products and services to the Australian community; Maquet Getinge Group for the generous donation of the ROTAFLOW centrifugal pump system to carry out the ex vivo ECMO model.

Funding

This work was supported by the National Health and Medical Research Council (1079421) CRE ACTIONS.

Author information

Authors and Affiliations

Authors

Contributions

JYS, MVM, JEM and JFF conceived the study and designed the experiment. KKK, MRP and CHHC performed the experiments and were responsible for data collection. MB performed statistical analysis. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Margaret R. Passmore.

Ethics declarations

Ethics approval and consent to participate

All methods were approved and performed in accordance with the guidelines and regulations set forth by the Metro North Ethics Committee (HREC/16/QPCH/320).

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ki, K.K., Passmore, M.R., Chan, C.H.H. et al. Low flow rate alters haemostatic parameters in an ex-vivo extracorporeal membrane oxygenation circuit. ICMx 7, 51 (2019). https://doi.org/10.1186/s40635-019-0264-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40635-019-0264-z

Keywords