In vivo testing of the low-flow CO2 removal application of a compact, platform respiratory device

Abstract

Background

Non-invasive and lung-protective ventilation techniques may improve outcomes for patients with an acute exacerbation of chronic obstructive pulmonary disease or moderate acute respiratory distress syndrome by reducing airway pressures. These less invasive techniques can fail due to hypercapnia and require transitioning patients to invasive mechanical ventilation. Extracorporeal CO2 removal devices remove CO2 independent of the lungs thereby controlling the hypercapnia and permitting non-invasive or lung-protective ventilation techniques. We are developing the Modular Extracorporeal Lung Assist System as a platform technology capable of providing three levels of respiratory assist: adult and pediatric full respiratory support and adult low-flow CO2 removal. The objective of this study was to evaluate the in vivo performance of our device to achieve low-flow CO2 removal.

Methods

The Modular Extracorporeal Lung Assist System was connected to 6 healthy sheep via a 15.5 Fr dual-lumen catheter placed in the external jugular vein. The animals were recovered and tethered within a pen while supported by the device for 7 days. The pump speed was set to achieve a targeted blood flow of 500 mL/min. The extracorporeal CO2 removal rate was measured daily at a sweep gas independent regime. Hematological parameters were measured pre-operatively and regularly throughout the study. Histopathological samples of the end organs were taken at the end of each study.

Results

All animals survived the surgery and generally tolerated the device well. One animal required early termination due to a pulmonary embolism. Intra-device thrombus formation occurred in a single animal due to improper anticoagulation. The average CO2 removal rate (normalized to an inlet pCO2 of 45 mmHg) was 75.6 ± 4.7 mL/min and did not significantly change over the course of the study (p > 0.05). No signs of consistent hemolysis or end organ damage were observed.

Conclusion

These in vivo results indicate positive performance of the Modular Extracorporeal Lung Assist System as a low-flow CO2 removal device.

Background

Mechanical ventilation (MV) is currently being used to normalize blood gas tensions in patients with acute respiratory distress syndrome (ARDS) or acute exacerbations of chronic obstructive pulmonary disease (ae-COPD). MV can further injure the lungs via ventilator-induced lung injury (VILI) [1, 2]. Reducing the airway pressures through non-invasive ventilation (NIV) or lung-protective ventilation techniques can improve patient outcomes; however, up to 25% of ae-COPD patients will fail NIV due to hypercapnia [3, 4]. The subsequent hypercapnia and acidosis require invasive MV to normalize blood gases [5, 6].

Extracorporeal CO2 removal (ECCO2R) devices provide a means to resolve the hypercapnia during lung-protective ventilation and in some instances allow extubation or NIV [7,8,9,10,11]. Typical ECCO2R duration for ae-COPD patients is 7 days or less [12]. The ongoing VENT-AVOID (NCT03255057) and REST (NCT02654327) randomized clinical trials aim to further expand the clinical evidence of using vv-ECCO2R as an alternative to invasive mechanical ventilation or in conjunction with low tidal volume ventilation, respectively.

We are developing the Modular Extracorporeal Lung Assist System (ModELAS) as a multi-functional respiratory assist technology capable of accommodating three levels of respiratory assist. The ModELAS is a highly compact, integrated pump-lung capable of being worn or integrated onto a wheeled console. The device can be configured to provide pediatric or adult complete respiratory assist (oxygenation and CO2 removal) or adult low-flow ECCO2R. Selection of the cannula and hollow fiber membrane (HFM) bundle size permits configuration for either pediatric or adult respiratory applications. The pediatric ModELAS utilizes a central cannulation and has been evaluated in acute animal studies [13, 14]. The adult complete respiratory assist application uses a dual lumen cannulation and is currently being evaluated in 30-day animal studies [15, 16].

In the present study, we evaluated the 7-day in vivo performance of the adult low-flow ECCO2R ModELAS in an awake ovine model. This is the first in vivo evaluation of the ModELAS configured for adult low-flow ECCO2R. This series of animal studies aims to validate benchtop CO2 removal and pump performance as well as evaluate sustained device CO2 removal, thrombogenicity of the device, and hemocompatibility.

Methods

The adult low-flow ECCO2R ModELAS utilizes a 0.65-m2 cylindrical, stacked, uncoated polymethylpentene HFM bundle (Fig. 1a). The benchtop results of our previously reported ECCO2R device utilized an identical bundle [17]. The primary difference between the previous and current version of the device are washout holes within the impeller. The targeted CO2 removal for this device is at least 35% of the metabolically produced CO2 in humans (~ 70 mL/min at normocapnia and at rest) at a blood flow of 500 mL/min.

Fig. 1
figure 1

Prototype of the adult low-flow ECCO2R ModELAS device showing the blood and gas flow pathways (a). All three respiratory assist applications utilize an identical pumping compartment. Cannulation and HFM bundle size differentiate each respiratory assist configuration. Photograph of a study animal wearing the holstered device (b)

The adult low-flow ECCO2R ModELAS was evaluated for 7 days in healthy, 41.2–54 kg Suffolk sheep (n = 6; 5 males, 1 female; 7–12 months of age) at the McGowan Institute for Regenerative Medicine’s Center for Preclinical Studies (CPCS). The animal care and surgical protocol was approved by the University of Pittsburgh’s Institutional Animal Care and Use Committee (Protocol# 17101264). All animals had unrestricted access to food and water and were able to stand and lie down. CPCS staff assessed the animals’ well-being, habitus, and pain regularly. Prior to the study, the animals were conditioned to the fixed-point tethering system and the modified backpack which held the device during the study.

The animals were fasted at least 12 h prior to the surgery. Anesthesia was induced via a subcutaneous injection of atropine (0.25 mg/kg) followed by intravenous administration of ketamine (4 mg/kg) and midazolam (0.4 mg/kg). The animals were then directly intubated using a cuffed McGill-type endotracheal tube and anesthesia was maintained with isoflurane.

Venous and arterial pressure lines were placed in the left external jugular vein and carotid artery, respectively. These lines were secured at the vessel, tunneled underneath the skin, and left in place for the entirety of the study. The right external jugular vein was exposed via a surgical cut down. The target pre-catheterization activated clotting time (ACT) was above 300 s (ACT-II, Medtronic, Minneapolis, MN) and was achieved with a 180 IU/kg heparin bolus. Additional 1000–2000 IU heparin boluses were administered to increase the ACT above 300 s if necessary. The right external jugular vein was cannulated with a 15.5 Fr dual-lumen catheter (ALung Technologies, Pittsburgh, PA). The catheter was advanced to the right atrial-superior vena cava junction. The catheter position was confirmed via fluoroscopy. The catheter was secured to the neck via sutures and further protected with a neck wrap.

The primed adult low-flow ECCO2R ModELAS (4 IU/mL heparinized normal saline) was connected to the catheter, and extracorporeal blood flow was initiated at 0.5 L/min. Pure oxygen sweep gas was initially set to 4 L/min. The animal was recovered from anesthesia in the CPCS intensive care unit and monitored 24 h a day. The animal wore the device in a modified backpack (Fig. 1b). The animal was tethered within a pen and able to freely stand, lay, and move its head during the study. Prophylactic antibiotics were given every 8 h (25 mg/kg cephazolin) and an analgesic (1 mg/kg flunixin meglumine) was administered as needed. Mean arterial pressure (MAP), central venous pressure (CVP), and heart rate were recorded hourly. ACT was measured at least every 8 h post-operatively and more frequently when heparin titration was required.

Prior to euthanasia, a heparin bolus (15,000 IU) was administered to increase the ACT to greater than 300 s. After confirmation of the ACT, the ModELAS pump was turned off and the device removed from the animal, gravity flushed with saline, and examined for thrombus. SomnaSol (pentobarbital sodium, Henry Schein Animal Health, Dublin, OH) was then administered, and cessation of the heartbeat was confirmed by the CPCS veterinarian.

CO2 removal rate (\( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \)) was measured daily and is reported as the average of 10 min of data collected at a sampling frequency of 2 Hz. Reported \( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \) was collected at a sufficiently high sweep gas flow rate (QSG) that the \( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \) was not a function of sweep gas flow rate [18]. The \( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \) was measured as the fraction of CO2 in the outlet sweep gas (\( {\mathrm{F}}_{{\mathrm{CO}}_2} \)) (WMA-4 Analyzer, PP Systems, Amesbury, MA). The \( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \) was calculated according to Eq. 1 and normalized to an inlet pCO2 of 45 mmHg according to Eq. 2 in order to reduce variability resulting from fluctuations of the inlet pCO2 [17]:

$$ {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2}={\mathrm{Q}}_{\mathrm{SG}}{\mathrm{F}}_{{\mathrm{CO}}_2} $$
(1)
$$ {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2}^{\ast }={\dot{\mathrm{V}}}_{{\mathrm{CO}}_2}\frac{45\mathrm{mmHg}}{{\mathrm{P}}_{{\mathrm{CO}}_2}^{\mathrm{Inlet}}} $$
(2)

The inlet pCO2 (\( {\mathrm{P}}_{{\mathrm{CO}}_2}^{\mathrm{Inlet}}\Big) \) used to normalize the \( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \) is the average of the device inlet pCO2 taken immediately before and after data collection (\( {\mathrm{P}}_{{\mathrm{CO}}_2}^{\mathrm{Inlet}} \)). During periods of low HCT (below 18%), device inlet pCO2 was only taken prior to data collection.

Plasma-free hemoglobin (PfHb), blood chemistry, and complete blood count (CBC) were measured pre-operatively and post-operatively on pre-operative days 6 and 1 and on POD (post-operative day) 0, 1, 3, 6, and 7. The blood sampling schedule was modified if the HCT was below 18% in order to reduce the amount of blood drawn from the animal. The reported pre-operative measurements are an average of data from both pre-operative days. POD 0 data is an average of samples taken when the arterial line was placed, initiation of extracorporeal blood flow and 4 h post-operatively. PfHb and platelet activation were measured using previously described methods [19,20,21]. Blood chemistry, manual platelet count, and CBC were measured by an external lab (IDEXX Laboratories, Inc., Westbrook, ME). Liver and kidney function and tissue injury were monitored by aspartate transferase (AST), alanine transferase (ALT), alkaline phosphatase (ALP), blood urea nitrogen (BUN), creatinine, and creatine kinase (CK). Platelet activation was measured via flow cytometry as previously reported [20, 21]. Platelet function was quantified by activating platelets with platelet activating factor (PAF).

Blood chemistry, platelet count, and CBC samples were not drawn on POD 6 for animals 3 and 6 due to low hematocrit. The blood chemistry for animal 4, pre-operative day 6 and manual platelet count for animal 6, preoperative day 1 and POD 1 and 3 were not measured due to a technical oversight.

Statistical analysis

Data points were averaged across studies for each POD and are reported as the average ± standard deviation. Statistical analysis was completed using IBM SPSS (IBM Corporation, North Castle, NY). Due to the instances of missing data previously described, the statistical analysis was completed using a mixed linear model analysis of variance and the restricted maximum likelihood estimation method [22]. Raw extracorporeal blood flow rate, motor torque, heart rate, CD62P expression, creatinine, CK, ALT, and BUN data violated the assumption of normality based on the Kolmogorov–Smirnov test and were normalized [23] prior to statistical evaluation. For those comparisons in which time had a significant effect, pairwise comparisons were conducted using least significant difference analysis. Statistical significance was considered at p < 0.05.

Results

All animals were successfully recovered from the surgery, and five studies were electively terminated on POD 7. Animal 4 was terminated early on POD 4 due to low MAP and an elevated lactate level. The necropsy showed significant pulmonary emboli within both lungs. Histopathological examination of the pulmonary emboli demonstrated that the initial emboli likely occurred several days prior to the end of study. On POD 1 of animal 4, the extracorporeal blood flow rate was 0.0 L/min for approximately 1 h. Blood pressure within the device was elevated, indicating an occlusion post-device. The device inflow tubing was clamped off while saline was forced through the device outflow tubing, and this returned blood flow to 0.5 L/min. Significant anemia was observed during animal 3 due to a non-study-related hematoma within the right scapular/axillary region. No thrombus or free fluid was evident within the chest or pericardium; thus, the scapular hematoma was likely due to accidental trauma to the shoulder unrelated to the device. Animal 5 was terminated approximately 3 h prior to the planned end of study due to a fractured positive pressure port on the device used to measure bundle inlet pressure on the prototype device. Significant blood loss through the fractured port necessitated rapid euthanasia, and POD 7 blood samples for blood chemistry, CBC, and gas exchange data were not collected.

Figure 2 provides average extracorporeal blood flow rate and \( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \). Average blood flow rate across all animals was 0.48 ± 0.01 L/min and did not vary significantly over the course of the 7 days (p = 0.338). Average CO2 removal (normalized to an inlet of 45 mmHg) across all animals and time points was 75.6 ± 4.7 mL/min. Daily average \( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \) did not vary significantly from POD 0 (p = 0.233). The average device inlet pCO2 was 43.6 ± 1.1 mmHg. The device inlet pCO2 values measured before and after the \( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \) data collection did not deviate by greater than 15%.

Fig. 2
figure 2

Raw and normalized CO2 removal rates (a) and daily extracorporeal blood flow rate (b). When compared to the POD 0 data, neither parameter significantly changed over duration of the study (p > 0.05). an = 5, bn = 4

MAP, CVP, and heart rate did not significantly change from POD 0 (Table 1). The negative CVP was likely due to the pressure monitoring line’s proximity to the 15.5 Fr DLC catheter drainage holes. Table 2 provides the pre-operative and post-operative animal hematologic parameters used to evaluate end organ function and hemocompatibility. Platelet function and activation did not statistically change from pre-operative values. BUN statistically increased on POD 6 prior to returning to pre-operative values on POD 7. HCT post-operatively was statistically less than pre-operative measurements. Post-operative pfHb remained below 25 mg/dL with the exception of POD 3–5 of animal 2 prior to returning to a normal level on POD 7 (Fig. 3). This animal showed signs of an acute kidney injury on POD 4 which resolved on POD 6. Macroscopic and microscopic examination showed no abnormalities or damage to the heart, lungs, liver, kidneys, or spleen, except animal 4 which exhibited signs of a pulmonary embolism.

Table 1 Hemodynamic and device parameters
Table 2 Hematologic and end organ function parameters
Fig. 3
figure 3

Plasma-free hemoglobin generated during the adult low-flow ECCO2R ModELAS in vivo study. Legend: animal 1; ■ animal 2; ▲ animal 3; animal trial 4; ♦ animal 5; animal 6

The explanted device from animal 2 had thrombus at the bottom pivot of the impeller and within the bundle (Fig. 4). During this animal study, the ACTs were below the target range for a quarter of the study. All other explanted devices were free of significant thrombus.

Fig. 4
figure 4

(a) Images of the explanted hollow fiber membrane bundle inlet (top row) and outlet (bottom row) faces and the (b) impeller top (top row) and bottom (bottom row) pivots. The animal 2 (right) bundle and pivots had significant thrombus formation due to low ACT. All other animals were free of significant thrombi (left)

Discussion

The Modular Extracorporeal Lung Assist System (ModELAS) is an integrated, modular pump-lung capable of providing three levels of respiratory assistance. Selection of the cannula and hollow fiber membrane bundle configures the ModELAS for pediatric complete respiratory assist, adult complete respiratory assist, or adult low-flow CO2 removal. The present study validated benchtop gas exchange performance and evaluated the thrombogenicity and hemocompatibility of the adult low-flow ECCO2R ModELAS in a healthy ovine model for 7 days. All animals were successfully recovered from the surgery, and a single animal required early termination on POD 4. No device exchanges were required.

These preclinical studies indicate that the adult low-flow ECCO2R ModELAS can operate for 7 days without diminished CO2 removal performance nor significant damage to blood or end organs. The average \( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \) was 75.6 mL/min, and the \( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \) on POD 7 was 83% of that on POD 0. Low HCT resulting from a shoulder hematoma (animal 3), pulmonary embolism (animal 4), and subcutaneous bleeding (animal 5) resulted in slightly diminished CO2 removal rates [24, 25] and increased blood flow rates during the latter half of the study. The in vivo CO2 removal performance on POD 7 was within 5% of previously published benchtop ModELAS \( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \) data [17, 25], hence validating benchtop performance.

The adult low-flow ECCO2R ModELAS was generally well tolerated by the animals. Platelet and organ function parameters used to evaluate device safety did not significantly change during the study (Table 2). Except animal 4, macroscopic and histopathological evaluation of the end organs did not indicate any device-related pathology. Pre-operative pfHb samples were drawn via a venous puncture and are therefore slightly elevated. A single study (animal 2) had elevated pfHb measurements greater than the 40–50 mg/dL clinically relevant threshold [26,27,28]. The maximum pfHb value occurred on POD 6 before decreasing on POD 7. The elevated pfHb was not associated with any hematuria. Animal 2 was also the sole animal in which the explanted device contained significant thrombus at the pivot bearing, likely caused by periods of improperly concentrated heparin infusion solutions. The thrombus and subsequent friction at the pivot bearing are possible contributors to the elevated pfHb. In addition, just prior to this period of elevated pfHb, the animal displayed symptoms of an acute kidney injury. The combination of the pivot thrombus and the kidney injury likely caused the sharp increase in pfHb in animal 2. Elevated pfHb was isolated to one out of the six studies, rather than a trend, and is not indicative of significant safety concerns.

The most significant device-related adverse event was a pulmonary embolism during animal 4. This animal required early termination on POD 4 due to low MAP and elevated heart rate and blood lactate concentration. The root cause of these symptoms was an acute pulmonary embolism in both lungs. This thrombus initiated several days prior to the end of study based on histopathological examination of the thrombus organization. Thromboembolism is a known risk of connecting any device to the vasculature including current artificial lungs [29, 30] and ventricular assist devices [31, 32]. In preclinical testing, the key is to differentiate whether the thromboembolism is inherent to the device design or was independent of the device used. The densely packed hollow fiber membrane bundle of the ModELAS acts as a filter; hence, it is unlikely that the thrombus was generated within the pump and migrated post-device (Fig. 1). Thus, the precipitating thrombus likely originated between the outlet of the device and the cannula tip, possibly at the tubing connectors [33,34,35], and was dislodged on POD 1 with the saline flush. The exact cause of the original thrombus remains unknown; however, it was an isolated event and does not appear to be characteristic of the device. The entire ModELAS circuit is also uncoated, whereas most commercial artificial lung circuits are coated with a biocompatible coating [36, 37]. Our laboratory is currently developing a tip-to-tip zwitterionic surface coating which will further improve hemocompatibility and further reduce thrombus formation within the ModELAS device, tubing, and catheter [38].

The most relevant comparison to the adult low-flow ModELAS is the Hemolung RAS (ALung Technologies, Pittsburgh, PA) which is currently undergoing an FDA clinical trial (VENT-AVIOD, NCT03255057) to evaluate the use of the Hemolung RAS in ae-COPD patients and the REST trial in the UK to assess the use of VV-ECCO2R in conjunction with the lung-protective ventilation in ARDS patients. In previously published pre-clinical healthy ovine studies, the Hemolung (0.59 m2) achieved 42–53 mL/min of CO2 removal at blood flow rates between 350 and 450 mL/min [27]. In comparison, the adult low-flow ECCO2R ModELAS CO2 removal rate ranged from 69 to 84 mL/min at blood flow rates ranging between 430 and 590 mL/min in the present study. The improved \( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \) in the adult low-flow ECCO2R ModELAS could be a result of slightly higher blood flow rates and bundle surface area (0.65 m2); however, the ModELAS has the additional benefits of being designed as a wearable unit and being capable of a wider range of respiratory therapies.

Conclusions

The ModELAS is a novel platform respiratory assist device. This preclinical study demonstrated that the low-flow ECCO2R ModELAS is capable of removing clinically significant amounts of CO2 at a blood flow rate of 500 mL/min consistently over 7 days in a healthy ovine model. The explanted devices were generally free of thrombus and only isolated device-related events occurred demonstrating positive hematologic and thrombotic compatibility of the ModELAS. The ModELAS is a promising platform respiratory assist technology, and further testing and product development is warranted.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

ACT:

Activated clotting time

ALT:

Alanine transferase

ALP:

Alkaline phosphatase

AST:

Aspartate aminotransferase

ARDS:

Acute respiratory distress syndrome

ae-COPD:

Acute exacerbations of chronic obstructive pulmonary disease

AST:

Aspartate transferase

BUN:

Blood urea nitrogen

CBC:

Complete blood count

CK:

Creatine kinase

CVP:

Central venous pressure

ECCO2R:

Extracorporeal CO2 removal

\( {\mathrm{F}}_{{\mathrm{CO}}_2} \) :

Fraction of CO2 in the sweep gas

HCT:

Hematocrit

MAP:

Mean arterial pressure

ModELAS:

Modular Extracorporeal Lung Assist System

MV:

Mechanical ventilation

NIV:

Noninvasive ventilation

PAF:

Platelet-activating factor

PAP:

Pulmonary arterial pressure

PfHb:

Plasma-free hemoglobin

POD:

Post-operative day

QSG :

Sweep gas flow rate

\( {\dot{\mathrm{V}}}_{{\mathrm{CO}}_2} \) :

CO2 removal rate

VILI:

Ventilator-induced lung injury

WBC:

White blood cell count

References

  1. Ricard J-D, Dreyfuss D, Saumon G (2003) Ventilator-induced lung injury. Eur Respir J 22:2s–9s

    Article  Google Scholar 

  2. Slutsky A, Ranieri V (2013) Ventilator-induced lung injury. N Engl J Med 369:2126–2136. https://doi.org/https://doi.org/10.1056/NEJMra1208707

  3. Plant PK, Owen JL, Elliott MW (2000) Early use of non-invasive ventilation for acute exacerbations of chronic obstructive pulmonary disease on general respiratory wards: a multicentre randomised controlled trial. Lancet 355:1931–1935

    Article  CAS  Google Scholar 

  4. Brochard L, Mancebo J, Wysocki M, et al (1995) Noninvasive ventilation for acute exacerbations of chronic obstructive pulmonary disease. N Engl J Med 333:817–822. https://doi.org/https://doi.org/10.1056/NEJM199509283331301

  5. Calverley PMA (2003) Respiratory failure in chronic obstructive pulmonary disease. Eur Respir J 22:26s–30s. https://doi.org/https://doi.org/10.1183/09031936.03.00030103

  6. Budweiser S, Jörres RA, Pfeifer M (2008) Treatment of respiratory failure in COPD. Int J Chron Obstruct Pulmon Dis 3:605–618

    Article  Google Scholar 

  7. Abrams D, Brenner K, Burkart K et al (2013) Pilot study of extracorporeal carbon dioxide removal to facilitate extubation and ambulation in exacerbations of chronic obstructive pulmonary disease. Ann Am Thoar Soc 10:307–314

    Article  CAS  Google Scholar 

  8. Bonin F, Sommerwerck U, Lund L, Teschler H (2013) Avoidance of intubation during acute exacerbation of chronic obstructive pulmonary disease for a lung transplant candidate using extracorporeal carbon dioxide removal with the Hemolung. J Thorac Cardiovasc Surg 145:e43–e44

    Article  Google Scholar 

  9. Kluge S, Braune S, Engel M et al (2012) Avoiding invasive mechanical ventilation by extracorporeal carbon dioxide removal in patients failing noninvasive ventilation. Intensive Care Med 38:1632–1639

    Article  Google Scholar 

  10. Bein T, Weber-Carstens S, Goldmann A et al (2013) Lower tidal volume strategy (~ 3 ml/kg) combined with extracorporeal CO2 removal versus ‘conventional’ protective ventilation (6 ml/kg) in severe ARDS. Intensive Care Med 39:847–856

    Article  Google Scholar 

  11. Terragni P, Del Sorbo L, Mascia L et al (2009) Tidal volume lower than 6 ml/kg enhances lung protection. Anesthesiology 111:826–835

    Article  Google Scholar 

  12. Trahanas J, Lynch W, Bartlett R (2017) Extracorporeal support for chronic obstructive pulmonary disease: a bright future. J Intensive Care Med 32:411–420. https://doi.org/https://doi.org/10.1177/0885066616663119

  13. May AG, Orizondo RA, Frankowski BJ, et al (2018) Acute in vivo evaluation of the Pittsburgh pediatric ambulatory lung. ASAIO J 65:395–400. https://doi.org/https://doi.org/10.1097/MAT.0000000000000918

  14. Orizondo RA, May AG, Madhani SP, et al (2018) In vitro characterization of the Pittsburgh pediatric ambulatory lung device. ASAIO J 64:806–811. https://doi.org/https://doi.org/10.1097/MAT.0000000000000711

  15. Madhani S, Frankowski B, Ye S et al (2017) In vivo 5 day animal studies of a compact, wearable pumping artificial lung. ASAIO J 65:94–100

    Article  Google Scholar 

  16. Madhani S, Frankowski B, Burgreen G et al (2017) In vitro and in vivo evaluation of a novel integrated wearable artificial lung. J Heart Lung Transplant 36:806–811

    Article  Google Scholar 

  17. May AG, Jeffries RG, Frankowski BJ, et al (2018) Bench validation of a compact low-flow CO2 removal device. Intensive Care Medicine Experimental 6:. https://doi.org/https://doi.org/10.1186/s40635-018-0200-7

  18. Federspiel W, Hattler B (1996) Sweep gas flowrate and CO2 exchange in artificial lungs. Artif Organs 20:1050–1052

    Article  CAS  Google Scholar 

  19. Svitek R, Frankowski B, Federspiel W (2007) Evaluation of a pumping assist lung that uses a rotating fiber bundle. ASAIO J 51:773–778

    Article  Google Scholar 

  20. Johnson CA, Wearden PD, Kocyildirim E, et al (2011) Platelet activation in ovines undergoing sham surgery or implant of the second generation PediaFlow pediatric ventricular assist device. Artif Organs 35:602–613. https://doi.org/https://doi.org/10.1111/j.1525-1594.2010.01124.x

  21. Johnson C Jr, Snyder T, Woolley J, Wagner W (2008) Flow cytometric assays for quantifying activated ovine platelets. Artif Organs 32:136–145

    Article  CAS  Google Scholar 

  22. Duricki DA, Soleman S, Moon LDF (2016) Analysis of longitudinal data from animals with missing values using SPSS. Nat Protoc 11:1112–1129. https://doi.org/https://doi.org/10.1038/nprot.2016.048

  23. Templeton GF (2011) A two-step approach for transforming continuous variables to normal: implications and recommendations for IS research. Commun Assoc Inf Syst 28:41–58. https://doi.org/https://doi.org/10.17705/1CAIS.02804

  24. Bidani A, Crandall ED (1982) Analysis of the effects of hematocrit on pulmonary CO2 transfer. J Appl Physiol Respir Environ Exerc Physiol 53:413–418. https://doi.org/https://doi.org/10.1152/jappl.1982.53.2.413

  25. May AG, Omecinski KS, Frankowski BJ, Federspiel WJ (2020) Effect of hematocrit on the CO2 removal rate of artificial lungs: ASAIO journal 1. https://doi.org/https://doi.org/10.1097/MAT.0000000000001140

  26. Extracorporeal Life Support Organization (2017) ELSO guidelines for cardiopulmonary extracorporeal life support, 5th ed. Ann Arbor, MI

  27. Wearden PD, Federspiel WJ, Morley SW, et al (2012) Respiratory dialysis with an active-mixing extracorporeal carbon dioxide removal system in a chronic sheep study. Intensive Care Med 38:1705–1711. https://doi.org/https://doi.org/10.1007/s00134-012-2651-8

  28. Omar HR, Mirsaeidi M, Socias S, et al (2015) Plasma free hemoglobin is an independent predictor of mortality among patients on extracorporeal membrane oxygenation support. PLoS One 10:e0124034. https://doi.org/https://doi.org/10.1371/journal.pone.0124034

  29. Dornia C, Philipp A, Bauer S, et al (2014) Analysis of thrombotic deposits in extracorporeal membrane oxygenators by multidetector computed tomography. ASAIO J 60:652–656. https://doi.org/https://doi.org/10.1097/MAT.0000000000000133

  30. Lubnow M, Philipp A, Foltan M, et al (2014) Technical complications during veno-venous extracorporeal membrane oxygenation and their relevance predicting a system-exchange—retrospective analysis of 265 cases. PLoS One 9:e112316. https://doi.org/https://doi.org/10.1371/journal.pone.0112316

  31. Starling RC, Moazami N, Silvestry SC, et al (2014) Unexpected abrupt increase in left ventricular assist device thrombosis. N Engl J Med 370:33–40. https://doi.org/https://doi.org/10.1056/NEJMoa1313385

  32. Uriel N, Han J, Morrison KA, et al (2014) Device thrombosis in HeartMate II continuous-flow left ventricular assist devices: a multifactorial phenomenon. J Heart Lung Transplant 33:51–59. https://doi.org/https://doi.org/10.1016/j.healun.2013.10.005

  33. Hastings S, Ku D, Wagoner S et al (2017) Sources of circuit thrombosis in pediatric extracorporeal membrane oxygenation. ASAIO J 63:86–92

    Article  CAS  Google Scholar 

  34. Matsuhashi Y, Sameshima K, Yamamoto Y, et al (2017) Investigation of the influence of fluid dynamics on thrombus growth at the interface between a connector and tube. J Artif Organs 20:293–302. https://doi.org/https://doi.org/10.1007/s10047-017-0973-6

  35. Griffith K, Jenkins E, Stulak J et al (2012) Long-term use of the CentriMag® ventricular assist system as a right ventricular assist device: a case report. Perfusion 27:65–70. https://doi.org/10.1177/0267659111424634

  36. Maquet Getinge Group (2015) HLS Set Advanced: integration meets innovation

  37. ALung Technologies, Inc (2012) Hemolung RAS: the first fully-integrated respiratory dialysis system

  38. Malkin A, Ye S, Lee E, et al (2018) Development of zwitterionic sulfobetaine block copolymer conjugation strategies for reduced platelet deposition in respiratory assist devices. J Biomed Mater Res B 106:2681–2692. https://doi.org/10.1002/jbm.b.34085

Download references

Acknowledgments

The authors thank the staff at the McGowan Institute’s Center for Preclinical Studies for their assistance with these studies.

Funding

This work was supported by the University of Pittsburgh’s Innovation Institute, the Coulter Translational Research Partners II Program at the University of Pittsburgh, and NIH grants R01HL117637 and R01HL135482. A.G.M. was supported by an NIH training grant (T32 HL076124) for the University of Pittsburgh’s Cardiovascular Bioengineering Training Program. R.A.O. was supported by the National Center for Advancing Translational Sciences of the NIH (TL1TR001858).

Author information

Authors and Affiliations

Authors

Contributions

All authors have contributed substantially in study conception, design, execution, and analysis. All authors have reviewed and approved the final manuscript draft and are accountable for all aspects of the work. Specifically, AGM, RAO, BJF, SY, EK, WRW, JD, and WJF in the conception and design; AGM, RAO, BJF, SY, EK, WRW, JD, and WJF in the analysis and interpretation; and AGM, RAO, BJF, SY, EK, WRW, JD, and WJF in the drafting of the manuscript for the important intellectual content.

Corresponding author

Correspondence to William J. Federspiel.

Ethics declarations

Ethics approval and consent to participate

The animal care and surgical protocol was approved by the University of Pittsburgh’s Institutional Animal Care and Use Committee. All animals received humane care.

Consent for publication

Not applicable

Competing interests

W.J.F. chairs the Scientific Advisory Board and is a founder of ALung Technologies, in which he has an equity interest. No other authors have conflicts of interest to report.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

May, A.G., Orizondo, R.A., Frankowski, B.J. et al. In vivo testing of the low-flow CO2 removal application of a compact, platform respiratory device. ICMx 8, 45 (2020). https://doi.org/10.1186/s40635-020-00329-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40635-020-00329-9

Keywords